The PrimeSight UNITY 9000 Video Processor User Manual
8 - Troubleshooting
52
PROBLEM
POSSIBLE CAUSE
SUGGESTED ACTION
Image quality Is
poor
Endoscope has not been
properly white-balanced to
the Processor
Perform another auto white balance.
Endoscope is not connected
properly
Ensure scope is properly connected to the
Processor.
Enhancement or other image
settings on the Processor are
not set correctly
Change the settings on the Processor.
The endoscope image is
being affected by another
device (e.g. ESU)
Make sure that the Processor and ESU devices
are connected to different circuits. An isolation
transformer may be required. Place electrical
equipment as far away from the endoscopy
equipment as possible.
Cables running from the
Processor to monitor are
damaged or improperly
connected
Inspect cables to ensure they are not damaged
and are properly connected. Cable may need
to be replaced or use an alternate video output
from Processor.
Endoscope is damaged
Send the endoscope and Processor to
LABORIE for repair.
Patient debris or other
material on the Objective
Lens
Clean the Distal Tip with an alcohol prep pad
to remove material or stain. Excess staining
may not be correctable or the lens may require
replacement.
Image has
blooming
The Brightness or Exposure
Mode controls may not be
set properly
Set the Exposure Mode to Auto and adjust
Brightness
The Video Processor is not
functioning properly
Return the Video Processor and endoscope to
LABORIE for repair.
Processor
or keyboard
locked up
Software glitch has
interrupted functions
Power the Video Processor off for 30 seconds,
then restart.