17
NOTE: Two fuses are used in this container, one is used in the unit and the other one
is supplied as a spare part. It’s recommended to replace the blown out fuse with the
fuse of same rating and type i.e. T2 5L 250
More information on replacing the fuses can be found in the maintenance section of this
guide.
•
Care instructions
o
Keep accessories away from dust when not in use, e.g. protect them from dust
cover.
o
Remove dust with a pneumatic rubber bulb and soft brush.
o
Use special optics cleaning cloths and pure alcohol for cleaning lenses and
eyepieces.
o
Protect your slit lamp from moist, fumes, acids and cosmetic materials. Do not
store chemicals close to the instrument.
o
Protect it from improper handling. Never install other devices sockets or unscrew
optical system and mechanical parts unless explicitly instructed to do so in this
manual.
o
Protect the microscope from oil and grease. Never oil or grease the guide
surfaces or mechanical parts.
o
Remove coarse contamination using a damp disposable cloth.
o
Use disinfectants based on the following ingredients: aldehydes, alcohols,
quaternary ammonium compounds.
o
Camera: Keep optical components clean using a lint-free cloth. Soak the cloth
using a little methanol or glass cleaner. Do not use ethanol and spirit.
o
Do not clean products with optical components in a cleaning/disinfecting device
or ultrasonic bath.
o
LABOMED MaxLite coatings are very resistant. If you clean as described
above, the coatings will not be damaged.