3 Installation
eVO500
/
eVO500
D Slit Lamp is shipped in a multi-layer foam container and it is divided
into major components: Microscope body, Binocular straight head, Base Platform, HD
Panel and Chin Rest Assembly, Illumination Column.
1.
Take out the various parts after removing
the wrapping and save the packing.
Fig. 1
2.
Mount the chin rest at the base platform
with the socket-head screws, hexagonal
wrench (provided) as shown in the Fig 2.
3.
Plug the wire from the chin rest assembly
into the 5 pin connector on the back of the
power supply unit.
Fig. 2
4.
Mount the base platform on appropriate
instrument stand arm or table base.
5.
Position the Slit Lamp body gently on the
base platform so that the geared rollers
match with the teeth on the guide rails.
When the instrument is pushed forward or
pulled backwards, the wheels should roll
freely and evenly over the guide-rails
(Refer Figure 3).
Fig. 3