background image

 
BTA 060, Rev 1K1 

Page 13/18 

6.4  Display menu (

diSP

The display menu for configuring the display contains the following items: 

Display 

Designation 

Description 

 

Display period for 
measured value (time 
data) 

Can be set between 0.5 and 99.9 s 

 

Display period for unit 
(time unit) 

Can be set between 0.0 and 99.9 s 

 

Rotate 180° 

Rotate screen by 180° when the device i s 

installed upside down 

 

Decimal places 

Setting the decimal places (zero to three 
fixed decimal places or automatic) 

 

Return 

Return to "

diSP

Table 8:  The items in the display menu 

6.4.1  Display period for measured value/unit (

td

 / 

tu

The  unit  can  be  displayed  in  two  ways,  either  by  pressing  the  middle  button  in  display 
mode  or  alternately  with  the  measured  value.  When  displayed  alternately  with  the 
measured value the display periods for the measured value and the unit can be selected 
independently of one another.  

If  the  period  for  displaying  the  unit  is  set  to  zero,  only  the  measured  value  will  be 
displayed.  

6.4.2  Rotating the display by 180° (

rot

You can rotate the 7-segment display by 180°, so th at it can be read when the device is 
put  in  place  upside  down.  The  function  for  the  arrow  buttons  are  also  swapped  in  this 
case, so that the device can be operated the same way in either position. 

Display 

Designation 

Description 

 

Standard (0°) 

 

 

Upside down (180°) 

Display rotated 180° for upside  down 

operation 

 

Return 

Return to "

rot

Table 9:  Parameter list for rotating the display 

Содержание PASCAL CS

Страница 1: ...Table of contents 1 General Information 2 1 1 Intended Use 2 1 2 General Safety Notes 2 1 3 CE Marking 2 2 Transportation and Storage 2 3 Installation and Commissioning 2 3 1 Electrical Connection 3 3...

Страница 2: ...ble EU Directives for placing products on the market within the European Union The following guideline applies to model series PASCAL CS21xx EMC Guideline EMC 2004 108 EG 2 Transportation and Storage...

Страница 3: ...switching output 2 Pgm1 2 programming pins Figure 1 Pin assignment for M12 circular connector device side 3 1 1 Connecting the switching outputs optional The switching outputs are potential free They...

Страница 4: ...nless steel housing with one hand and turn the display unit with the other hand into the wanted position The turning angle is limited by an internal limit stop Do not try to force the display unit bey...

Страница 5: ...configure it with a communication modem and the PC Software COMLINE S This is described in a separate manual 4 2 Maintenance When properly installed in accordance with applicable specifications this...

Страница 6: ...y the LED flashing When there is no button pressed the output states are displayed 6 1 2 Display mode Measured value screen When the device is switched on it goes into display mode The currently measu...

Страница 7: ...to display mode At the end of a menu typically the rEt item you return to the first menu item by repressing the down arrow button again Similarly you can jump from the first menu item to the end of th...

Страница 8: ...the rEt item which allows you to return to the next higher menu level The middle button stores your selection Stor appears on the display to confirm that the value has been stored and the device retur...

Страница 9: ...the previous value You confirm the selected digit with the middle button and proceed to the next digit You can cancel the value entry at any time by pressing both arrow buttons simultaneously The dev...

Страница 10: ...system data versions serial number reset to factory settings Return Return to display mode Table 3 The items in the main menu 6 3 Basic menu bASE The basic menu contains the following items Display D...

Страница 11: ...of the measuring range with the teach lower range menu item tlr When the menu item is selected the entry points to rEt To trigger the function go to YES with one of the arrow buttons and confirm with...

Страница 12: ...m in this menu Resetting a pointer is confirmed by showing on the display Display Designation Description Value of min max pointer Value of min max pointer in the selected measuring unit Clear Reset t...

Страница 13: ...two ways either by pressing the middle button in display mode or alternately with the measured value When displayed alternately with the measured value the display periods for the measured value and...

Страница 14: ...shown on the display will appear flashing e g 99 99 when two decimal places are set for a measured value of 110 mbar 6 5 Switch point menu The switch point menu contains the functions for setting the...

Страница 15: ...hen you select a frame function the menu items for switch point and reset point are replaced by the upper and lower frame limits The minimum difference of the frame limits is also 0 5 of the measuring...

Страница 16: ...If the pressure is above the switch point the switch is open At the lower range limit the switch is closed Frame normally open Inside of the frame the switch is closed At the lower range limit the swi...

Страница 17: ...enu Due to the limited number of alphanumerical segments on the display hardware and software versions are split into two separate items and serial number into four items 6 6 2 Reset to factory settin...

Страница 18: ...Display direction 0 normal 180 turned Setting the decimal places zero to three fixed decimal places or automatic Configuring the switching outputs only if available Switch point or upper frame limit o...

Отзывы: