  BA_067_2019-05_10.00  Inline resistance thermometer 

Page 4/4 

 

Operation 

During operation, take care that the device remains within its intended pressure and tem-
perature ranges. No other monitoring is necessary. 

  

Permissible media temperature:  

-

20…200 °C 

  

Permissible ambient temperature:  

-4

0…100 °C 

When supplied with a transmitter, the permissible ambient temperature is reduced to 85°C. 

  

Restricted  ambient  conditions  may  apply  in  explosion  protected  environments  (see 
XA_004 for GA2201 or XA_002 for GA2211). 

  

4.1  Maintenance / Service 

When  properly  installed  in  accordance  with  applicable  specifications,  this  device  is 
maintenance-free. However, we recommend an annual recalibration of the device. 

  

In the event of a device malfunction, you can only replace the measuring insert respective-
ly the head transmitter. 

  

Disassembly 

When measuring hot media, make sure that the device has cooled down prior to any dis-
mounting or wear appropriate protective clothing to avoid burns. 

  

Switch off the power supply to the device before disconnecting the electrical connections. 
Once this is done, the device may be mechanically removed. 

  

 

Warning 

Opening pressurized lines might cause severe injuries. 

Danger of severe injuries or damage 

  Relieve  the  process  pressure  before  attempting  to  remove  the  device. 

Shut off the pressure supply for all feed lines to the device and relieve the 
pressure in them. 

  

 

Warning 

Hazardous deposits and  residues might remain on opened process connec-
tions and removed devices. 

Danger of injury 

  After  the  device  has  been  removed,  seal  off  the  measuring  point  and 

mark the open process connection accordingly. Consider a possible dan-
ger due to residues when handling the removed device. 

  