BA_080_en_2022-01_11.02  Pressure transmitter PASCAL CV4 

Page 14/24 

 

6.4.3 

Menue "Adjustment" 

The following functions are available for the pressure adjustment: 

  

Menue entry

 

Description

 

Zero point 

Set device at ambient pressure to zero bar 

(not for absolute pressure devices)

 

Position correct. 

Correct zero point error due to installation position  

(not for absolute pressure devices)

 

Lower adjust. 

Offset correction with applied reference pressure 

Upper adjust. 

Span correction with applied reference pressure 

Table 8: "Pressure Adjustment" menue 

  

Additionally, you can perform a current adjustment (see 6.4.3.4). 

  

6.4.3.1 

Zero-point correction 

The zero-point correction requires a non-pressurized pressure port e.g. during calibration in 
the lab. When executing this function the applied pressure will be interpreted as zero bar 
pressure. 

The zero point correction results in an offset of the whole characteristic curve. Therefore it 
is a special case of the lower adjustment (see below). 

  

6.4.3.2 

Position correction 

Using the function "Position correction", you can correct the offset error due to the installa-
tion position independent of the zero-point offset.  

In the corresponding submenue, you will find the following entries: 

  

Menue entry

 

Description

 

Pos.corr. on/off 

Activate/deactivate position correction 

Pos.corr. adjust. 

Correct position error at ambient pressure 

Table 9: "Position Correction" submenue 

  

When you set the position correction, it will be activated automatically. You can deactivate 
the  position  correction,  for  instance  for  a  subsequent  calibration  to  check  the  zero  point 
independently from the installation position. 

  

 

 

Содержание CV4110

Страница 1: ...ity with EU Regulations 2 1 4 EX Approval 2 2 Transportation and Storage 2 3 Installation and Commissioning 2 3 1 Devices with Diaphragm Seal 3 3 2 Differential pressure devices 3 3 3 Mechanical Insta...

Страница 2: ...nded to measure pressure of gases vapors and liquids as specified in the data sheet 1 3 Conformity with EU Regulations The CE marking on the device certifies its compliance with the applicable EU Dire...

Страница 3: ...l measurements it is the lower process connection 3 3 Mechanical Installation Use gaskets if required that are suitable for the process connection and resistant to the media Before starting operation...

Страница 4: ...ock completely downwards before inserting the cable Otherwise it is not possible to ensure a safe electrical connection Do not forget to tighten the cable gland after the electrical connection is fina...

Страница 5: ...rmed in the factory 5 Disassembly When measuring hot media make sure that the device has cooled down prior to any dis mounting or wear appropriate protective clothing to avoid burns Switch off the pow...

Страница 6: ...device The gen eral functionality of the buttons is identical in all operating modes Figure 2 Control elements If you press and hold the up or down button it will automatically be triggered multiple t...

Страница 7: ...value is larger than the displayable number due to setting a fixed decimal point or an inappropriate unit the largest displayable number is shown flashing In this case please select a different setti...

Страница 8: ...d by the header text Menu locked in combination with the lock icon Use the same key combination to unlock the menue again 6 2 3 Icons indicating device status In the top right corner of the display de...

Страница 9: ...lue is shown at the maxi mum size best readability from a dis tance One additional value can also be shown Table 4 Display layouts You can define what information shall be shown in the layout s placeh...

Страница 10: ...by pressing and holding the ESC button for at least one second Figure 4 Navigating the operating menue In long menues and selection lists a scroll bar on the left side shows the position of the curren...

Страница 11: ...and go back to view mode The se lected value is not saved With the OK button the selected value is saved Saving the value is confirmed with an in formation window and the new value is shown in view m...

Страница 12: ...then set to the parameter limit You can set negative numbers by decreasing the value below zero When you press the ESC button you can cancel the entry at any time and return to the display of the set...

Страница 13: ...ystem Device data and function such as factory reset Table 6 The main menue 6 4 2 Menue Quick Setup In Quick Setup basic configuration options are combined to make it possible to quickly con figure ke...

Страница 14: ...calibration in the lab When executing this function the applied pressure will be interpreted as zero bar pressure The zero point correction results in an offset of the whole characteristic curve Ther...

Страница 15: ...ustment should be performed as close as possible to the lower range value and upper range value of the measurement ranges 6 4 3 4 Current adjustment You can use the current adjustment if the reading a...

Страница 16: ...reen mode Configuration of the screen layout and content see 6 2 4 Decimal point Select setting of decimal point to determine number of decimal places Backlight Switch backlight on off Table 10 Displa...

Страница 17: ...psi Pound force per square inch 1 psi 68 9 mbar atm Atmospheric pressure 1 atm 1013 mbar mmH2O Millimetre water column 1 mmH2O 0 0981 mbar mH2O Metre water column 1 mH2O 98 1 mbar inH2O Inch water co...

Страница 18: ...tput Configure the measured value as well as the current output in the Measurement output menue Menue entry Description Damping Setting the damping of the output signal Output function Setting the out...

Страница 19: ...2 Output function You can set the output function proportionally rising to the measured value selection Lin ear or proportionally falling selection Inverse You can also define the output signal by a...

Страница 20: ...pressure For instance the 20 point may not be before the 10 point Otherwise you cannot activate the table function If the first table point is not set at 0 the start of the table is calculated by mean...

Страница 21: ...nt simulation influences only the current output The pressure simulation takes all settings into consideration including the damping 6 4 8 Menue Communication In the Communication menue the settings f...

Страница 22: ...settings Device reset Restarting the device such as after a power outage Table 18 System menue 6 4 9 1 Device ID Using the device ID you can show a custom text in the display if you configure the scre...

Страница 23: ...BA_080_en_2022 01_11 02 Pressure transmitter PASCAL CV4 Page 23 24...

Страница 24: ...version 1 1 0 and above Setting of the pressure value that should correspondto 20 mA endof range Setting the damping of the output signal Selection of the decimal places of the 1st value by specifyin...

Отзывы: