4.2 

Specific considerations 

To ensure that the build in safety features of the cabinets function properly, the location of installation 
must be selected carefully. The recommended ambient conditions, for safe and proper operation, 
are listed below. 
 

 

All cabinets are developed and produced for use in clean environments. 
 

 

Cabinets must not be operated outdoors or in environments with extreme air pollution. The 
safety cabinets are not intended to filter vapours containing acids or organic solvents. The 
safety cabinets must not be used as a fume hood. 

 

 

Up to 2000 m above sea level. 
 

 

The wall socket to which the cabinet is connected must be fused by a minimum 10 A fuse. 
 

 

The wall socket should be placed so accidental switch off is unlikely. A wall socket without 
a switch is recommended.   
 

 

The power cord is the mains disconnect. 
 

 

The location must be capable of supporting the weight of the device.  
 

 

If the factory-approved stand is not purchased, the holding device must be rated for twice 
the cabinet weight.  
 

 

Adequate room height. For cabinets, which are not connected to an exhaust system, the 
distance between the exhaust air opening on the cabinet and the room ceiling must be at 
least 400 mm.  

 

 

There must be at least 100 mm clearance on the rear and non-utility side of the cabinet. 

 

 

For sides with external utility connections leave 150 mm clearance. 
 

 

The location must be equipped with an appropriate ventilation system.  
 

 

Room temperature between 10 °C and 35 °C.  
 

 

Relative humidity must not exceed 80%, up to 31 °C and is decreasing linearly to 50% at 
40 °C.  
 

 

The cabinets are for indoor use only. 

 
Room  ventilation  should  preferably  be  a  ventilation  system  that  complies  with  the  national 
requirements for the application. 
 
The inlet air and exhaust air openings of the room ventilation must be located so that drafts do not 
impair the function of the sterile cabinet air system. 

4.3 

Transport  

The cabinet can be moved through a standard 800 mm wide door. 
 
Transport  of  the  Fortuna  cabinets  can  be  carried  out  by  lifting  the  cabinet  using  a  forklift  either 
sideways  under  the  support  stand  or  directly  under  the  trough.  Furthermore,  the  cabinet  can  be 
moved manually by using hooks in the dedicated holes in the support stand. 

Содержание Fortuna 1200

Страница 1: ...12162017QA01 Item No 700745 CopyRight 2017 Page 1 of 49 INSTRUCTION MANUAL Fortuna 900 1200 1500 1800 Clean Bench...

Страница 2: ...nger of medium or minor injuries or physical damage This document is the original instruction manual in accordance with the current Machinery Directive This document is the reference for any translati...

Страница 3: ...12 4 6 Connecting utilities to the Fortuna cabinets 12 5 Start up of the Fortuna cabinets 13 5 1 Preparation 13 5 2 Connections 13 5 3 Assemble the stand for the Fortuna cabinets 14 5 4 Levelling of...

Страница 4: ...27 8 3 Optional UV lights 28 8 4 Replacement of fuses 28 8 5 Replacement of laminator 29 8 6 Replacements and repairs 29 8 7 Disposal of cabinet 29 9 Specifications 30 9 1 Spare parts 34 10 Quality t...

Страница 5: ...factory settings The EN 12469 standard requires that the downflow air velocity is in the interval 0 25 m s 0 50 m s The LaboGene default downflow setting is 0 32 m s for 350 mm work opening with a max...

Страница 6: ...e nameplate data Caution The proper function and safety of the cabinet is only secured if personnel authorized by us perform the required tests maintenance and repair work Please also refer to section...

Страница 7: ...ing of downflow air velocity guarantees that product protection remains uncompromised Airflow monitoring using build in flow sensor determine the velocity of the airflow in the work chamber As soon as...

Страница 8: ...ired for individual requirements See Service menu When switching to reduced velocity the low flow alarm is activated and the light in the cabinet is turned off to indicate not safe conditions 3 Button...

Страница 9: ...w button 9 On start up after power disconnect or after power failure the above alarm will be active Press the blue ENTER button to clear the alarm The main screen showing the time and date The display...

Страница 10: ...ng chamber The standard LaboGene control system is prepared for full control of safety and timing for the use of UV light see section UV light option for operation and build in safety measures Remembe...

Страница 11: ...must be rated for twice the cabinet weight Adequate room height For cabinets which are not connected to an exhaust system the distance between the exhaust air opening on the cabinet and the room ceili...

Страница 12: ...f moving is warranted both the Fortuna can be lifted and moved using a low lifter on 1500 and 1800 models use one from each side Be aware that the cabinets are top heavy Stabilize cabinet during the m...

Страница 13: ...ront window manually this will cause window to mall function 3 Check that the window slides from top to bottom without any irregular sound 4 If the window does not move when the UP or DOWN button is p...

Страница 14: ...and Fortuna 1800 75 80 cm 9 001 040 014 Stand Fortuna 900 90 95 cm 9 001 030 011 Stand Fortuna 1200 90 95cm 9 001 030 012 Stand Fortuna 1500 90 95cm 9 001 030 013 Stand Fortuna 1800 90 95 cm 9 001 030...

Страница 15: ...net General functionality is checked ad Hoc To ensure optimal function and performance of the cabinet the parameter values found in the Factory test report must be used as close as possible An install...

Страница 16: ...be adjusted to any level desired for individual requirements See Service menu Button 3 with red light to silence the acoustic alarm Button 4 with blue light for normal light ON OFF Button 5 with yello...

Страница 17: ...Auto start auto stop Here it is possible to select Monday to Friday whole week or only a single day when the cabinet should start up in full speed and after how long it should wait until it switches o...

Страница 18: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 18 of 49 6 1 Overview of the user menu...

Страница 19: ...tantly supervised so that any deviation from safe conditions or any fault in the supervision system will be indicated immediately Safe conditions are indicated by a green LED light in the button for f...

Страница 20: ...mode if there is an alarm the light will turn off automatically and will only be turned on again when all alarms have been deactivated 6 6 Start up of the Mars and Mars Pro cabinets Turn on the cabine...

Страница 21: ...nly be activated when the window is completely closed The light is switched off in the cabinet during the UV disinfection period The UV period is controlled by a timer integrated in the control system...

Страница 22: ...front window is positioned in working position and kept in that position during the entire work process Necessary appliances for use during work must be placed within easy reach Secure the appropriate...

Страница 23: ...he work chamber remove fluids if any from the trough and dry it Disinfect if required Leave the cabinet fan in operation for about another 10 minutes Close the front window to its lowest position 6 11...

Страница 24: ...re for this process Decontamination using formaldehyde can be performed if a sterile working chamber is required for the work procedure This decontamination procedure or a similar one is required befo...

Страница 25: ...ring pads or steel wool If the coated surfaces do become dull streaked smeared or marred in some other way there is no known method to restore the finish 7 3 Disinfection of working chamber 1 Remove a...

Страница 26: ...on Before repair work is carried out inside the contaminated and or infected cabinet and prior to filter replacement and upon change of location proper cleaning and disinfecting by the operator is req...

Страница 27: ...onditions of the cabinet Optional test can be performed on request Particle test in the working chamber to determine the air quality Sound level test Light intensity test normal light Light intensity...

Страница 28: ...se lamps from the fixtures by rotating it 90 Install the new UV lamp performing the inverse operation Note Use gloves when manipulating a germicidal UV lamp to prevent leaving hand prints that reduce...

Страница 29: ...the replacement laminator and mount the fixing screws again Product name Catalogue No Laminator Fortuna 900 9 000 071 001 Laminator Fortuna 1200 9 000 071 002 Laminator Fortuna 1500 9 000 071 003 Lam...

Страница 30: ...at least Class H14 Table top and trough material AISI 304 stainless steel Table top max total load kg 100 Side glass material 5 mm hardened glass Front glass material 6 mm laminated safety glass Cabin...

Страница 31: ...m hardened glass Front glass material 6 mm laminated safety glass Cabinet and back wall material colour Polyester coated steel RAL 9016 Dimensions External dimensions with adjustable Stand Standard Dx...

Страница 32: ...hardened glass Front glass material 6 mm laminated safety glass Cabinet and back wall material colour Polyester coated steel RAL 9016 Dimensions External dimensions with adjustable Stand Standard DxW...

Страница 33: ...hardened glass Front glass material 6 mm laminated safety glass Cabinet and back wall material colour Polyester coated steel RAL 9016 Dimensions External dimensions with adjustable Stand Standard DxW...

Страница 34: ...0 9 000 050 008 Light ballast for Fortuna 1500 and 1800 9 000 050 009 Window motor for Fortuna 9 000 050 112 Display board Fortuna 9 000 050 011 Motor board Fortuna 9 000 050 010 UV light ballast Fort...

Страница 35: ...be tested Test Factory Installation Service Leak test of HEPA filters Yes Yes Yes Downflow air velocity and setting of alarms Yes Yes Yes Test of alarms Yes Yes Yes Light test Yes Yes Airflow pattern...

Страница 36: ...led the operator doing the work and the environment in the surrounding room All tests should be done by authorized test technician approved by LaboGene 11 Options for Fortuna cabinets A number of opti...

Страница 37: ...onal valves for utilities are delivered make sure that local rules and regulations are complied with during installation and use It is generally recommended to use an authorized technician for this ki...

Страница 38: ...12162017QA01 Item No 700745 CopyRight 2017 Page 38 of 49 12 Annex 1 Factory check list and factory test report...

Страница 39: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 39 of 49...

Страница 40: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 40 of 49...

Страница 41: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 41 of 49...

Страница 42: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 42 of 49...

Страница 43: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 43 of 49...

Страница 44: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 44 of 49...

Страница 45: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 45 of 49...

Страница 46: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 46 of 49...

Страница 47: ...o 700745 CopyRight 2017 Page 47 of 49 13 Annex 2 Service log book Type Product name Serial no Supplier Service interval Location Service repair Fault occurred on Notified date Repaired date Remark fau...

Страница 48: ...es from all dangers We therefore kindly ask you to copy and complete this statement before work is commenced Subject Service report no Service agreement no The undersigned hereby declares that the abo...

Страница 49: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 49 of 49 15 Annex 4 Declaration of comformity Fortuna...

Отзывы: