3
3. Installation or rotors and rotor maintenance
3.1 Rotors and accessories
The following accessories are available for the Spectrafuge:
Angle rotor for 6 x 15ml tubes
Order
no.
Included with unit (C0060-RTR)
Tube measurement
15ml (17x120mm)-10ml (16x100mm)
Max. speed
6,500rpm
Centrifuging radius
8.5cm
RCF (g-value)
4,000 x g
Adapter for 5ml(12x75mm) and 7ml (13x100mm) tubes
Order
no.
C0200-17A
Tube measurement
12x75mm, 13x100mm, or common
Sarstetd style tubes
Tube and tube adapter reference chart.
3.2 Rotor maintenance
The rotor should be cleaned thoroughly after each use.
Thorough
cleaning must be performed when spinning samples containing
phenol or phenol chloroform.
Periodically inspect the rotor for dents,
dings, scratches, discoloration and cracks If any damage to the rotor is
found, discontinue use of the rotor immediately and replace.
3.3 Removing and Installing the angle rotor
Remove the rotor screw from the motor shaft by turning the screw
counterclockwise. Lift the rotor upward and remove from the centrifuge.
Содержание Spectrafuge 6C
Страница 1: ...Lit M00031 Rev 2 August 2016 SPECTRAFUGE 6C User Manual C0060 C0060 230V...