5.  Service and Maintenance 

5.1  Centrifuge service 

The Spectrafuge 6C requires no routine maintenance other than the 
occasional routine cleaning.  All repairs should be performed by authorized, 
qualified personnel only.  Repairs performed by unauthorized personnel 
may void the warranty. 

5.2  Cleaning the centrifuge 

Always keep the centrifuge housing, rotor chamber, rotor and rotor accessories         
clean.  All parts should be wiped down periodically with a soft cloth.  For more      
thorough cleaning, use a neutral cleaning agent (pH between 6 and 8) applied 

           with a soft cloth.   

Excessive amounts of liquid should be avoided.  Liquid should not come into                     
contact with the motor.  After cleaning, ensure that all parts are dried thoroughly  
by hand or in a warm air cabinet (maximum temperature 50ºC) 

 

5.3  Cleaning the rotor 

The rotor should be cleaned after each use.  When spinning samples 
containing phenol or phenol chloroform, the rotor should be cleaned 
immediately after use.

 

5.4  Disinfection 

Should a spill of infectious materials occur within the rotor or chamber, the 
unit should be disinfected.  This should be performed by qualified personnel 
with proper protective equipment. 

5.5  Replacing fuses 

Check the fuse when it is recommended in the Troubleshooting Guide 
located in this manual.  The fuse holder is located in the power inlet on the 
rear of the unit.  Disconnect the power cord from the power inlet.  Open the 
fuse holder drawer by inserting a small screwdriver under the tab and prying 
it open.  Remove the innermost (operative) fuse from its retaining tabs and 
replace the fuse if necessary. A spare fuse is located in the outermost 
chamber of the fuse drawer.  Replace only with a fuse of exactly the same 
value as the original.  (Fuse type may be found in the Technical data section 
of this manual.) 

6.  Troubleshooting Guide 

Please refer to this guide before calling for service. 

Centrifuge will not start 

Possible reason:  

No power supply 

Solution: 

 

 

Check that power is being supplied to the outlet 
Check that the power cord is plugged into both 
the wall outlet and the back of the centrifuge.                      

 

 

 

 

Check that power cord is not damaged 

Possible reason:  

Blown fuse 

Solution: 

 

 

Check fuse and replace if necessary 

Содержание Spectrafuge 6C

Страница 1: ...Lit M00031 Rev 2 August 2016 SPECTRAFUGE 6C User Manual C0060 C0060 230V...

Страница 2: ...is designed to assist you in the optimal usage of your Labnet Spectrafuge 6C Centrifuge The manual is available in English French German Italian Portuguese and Spanish on our website at www labnetinte...

Страница 3: ...same type and weight ALWAYS locate the centrifuge within easy access to an electrical outlet ALWAYS use only centrifuge tubes designed to withstand centrifugal forces of at least 4 000 xg ALWAYS use a...

Страница 4: ...3 Installing the Rotor 2 Rotors and accessories Rotor maintenance Removing and installing the angle rotor Overloading rotors 4 Operation 5 Closing the lid Lid release Lid lock Speed selection Selecti...

Страница 5: ...any of the following conditions exist The centrifuge has not been installed properly The centrifuge is partially dismantled Service has been attempted by unauthorized or unqualified personnel The rot...

Страница 6: ...m the carton Retain the carton and cushions until it has been established that the centrifuge is working properly Inspect the centrifuge for any visible signs of shipping damage Shipping damage is the...

Страница 7: ...5mm 13x100mm or common Sarstetd style tubes Tube and tube adapter reference chart 3 2 Rotor maintenance The rotor should be cleaned thoroughly after each use Thorough cleaning must be performed when s...

Страница 8: ...ng page When loading the rotor refer to figure 3 located on the following page Loading in the pattern indicated will ensure a balanced load Tubes to be loaded should be filled equally by eye The diffe...

Страница 9: ...th the following formula Reduced speed nred x max speed nmax Example Where the density of the liquid is 1 7 the new maximum speed would be calculated as follows nred x 6 500 5 461 rpm If in doubt conc...

Страница 10: ...00 is displayed 4 4 Speed selection The speed rpm can be selected from 300 to 6 500rpm with the knob right The set speed can be viewed at all times on the large LED display right 4 5 Selection of oper...

Страница 11: ...infectious materials occur within the rotor or chamber the unit should be disinfected This should be performed by qualified personnel with proper protective equipment 5 5 Replacing fuses Check the fus...

Страница 12: ...ed from 8 30am to 5 30pm EST Monday through Friday Our 24 hour fax number is 732 417 1750 Inquiries may also be sent via our electronic mailbox at labnetinfo corning com Should your Spectrafuge requir...

Страница 13: ...9...

Страница 14: ...and electronic equipment WEEE Labnet Spectrafuge 6C is marked with the crossed out wheeled bin and must not be disposed of with domestic waste Consequently the buyer shall follow the instructions for...

Страница 15: ...aint or finish Claims for transit damage should be filed with the transportation carrier ALL WARRANTIES INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMIT...

Страница 16: ...9299870000 31 Mayfield Ave Edison NJ 08837 P 732 417 0700 F 732 417 1750 Toll Free 888 522 6381 labnetinfo corning com www labnetinternational com...

Отзывы: