Labnet System
Labnet Enduro™ GDS II
Gel Documentation System
Labnet Enduro GDS Touch II
Gel Documentation System
Camera
2.7 MP image resolution
2.7 MP image resolution
Lens
f/1.8
f/1.8
Light Sources
Epi White
302 or 365 nm Transilluminator
Epi White
302 or 365 nm Transilluminator
Epi Blue
Trans White Light
Illumination
Tray
Optional
Optional
Field of View
15 x 20 cm
15 x 20 cm
Cabinet
Door sensor to prevent accidental exposure
to UV; automatically shuts off UV light when
door is open
Safety override, enabling band cutting on the
UV transilluminator with automatic turn off
after 5 minutes
Door sensor to prevent accidental exposure to UV;
automatically shuts off UV light when door is open
Safety override, enabling band cutting on the UV
transilluminator with automatic turn off after 5
minutes
Emission Filter
590 nm
590 nm
3-position Filter Slider
No
Yes
Certifications
CE, cTUVus
CE, cTUVus
Product Footprint
12” x 15” (30.48 x 38.1 cm)
12” x 15” (30.48 x 38.1 cm)
Computer
Operating system: Windows® 7 1 gigahertz
(GHz) or faster 32-bit (x86) or 64-bit (x64)
processor (Intel), 2 gigabyte (GB) RAM
(32-bit) or 4 GB RAM (64-bit), 16 GB available
hard disk space (32-bit) or 20 GB (64-bit)
Operating system: Windows® 10 Processor:
1 gigahertz (GHz) or faster processor (Intel),
RAM: 2 gigabyte (GB) for 32-bit or 4 GB for
64-bit, Hard disk space: 16 GB for 32-bit OS
20 GB for 64-bit OS
Minimum hardware requirements: 1.4 GHz
processor speed, 2 GB RAM, 16 GB free hard
disk space, 2 USBs (camera and printer)
Tablet included: Windows 8 tablet computer enables
compatibility with existing networks, as well as easy
setup
Bluetooth and wireless compatible, or USB to LAN
compatible
10.1” touch screen
2 GB of storage for short-term data saving
3. Specifications
4. System Placement
Please make sure the system is located away from regular exposure to water, solvents, or corrosive material. Place on a flat and stable
surface with 10 cm clearance on all sides, including the top, to allow sufficient air flow.
The system is intended for indoor use with the following ambient conditions:
a. Altitude up to 2,000 m
b. Temperature 5°C to 40°C
c. Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C
The system should be placed away from interfering electrical signals and magnetic fields. If possible, a dedicated electrical outlet should be
used to eliminate electrical interference from other instrumentation in your laboratory.