Maintenance conditions
START-UP
Prior to switching the centrifuge on, one shall read carefully all sections of
this instruction in order to ensure smooth operation and avoid damages of
this device or its accessories.
In order to protect the centrifuge against unbalance, fill in the test tubes up
to the same weight.
TRANSPORTATION
Centrifuge must not be transported with the rotor mounted on the shaft.
GENERAL HINTS
One must use original rotors, test-tubes and spare parts only.
In case of faulty operation of the centrifuge one shall ask for assistance
service of LABNET INTERNATIONAL/Corning or its authorized
representatives.
NEVER switch the centrifuge on if it is not installed properly or rotor is not
fitted correctly.
CENTRIFUGING SUBSTANCES
NEVER exceed load limit set by the manufacturer. Rotors are intended for fluids
of average homogeneous density equal to 1,2 g/cm
3
or smaller when
centrifugation is carried out at maximum speed. When fluids of higher density
shall be used, then it is necessary to change density of centrifuges sample in
PARA/DENSITY field.