© 2018 Corning Incorporated
36
Automatic lid open
Automatic lid open
OPEN LID AFTER RUN
Via
keys choose AUTOM. LID
OPENING.
Press SET (to switch off/on).
When centrifuge process is finished, cover
will be opened automatically.
When centrifuging is terminated by
pressing STOP, opening cover is possible
by pressing COVER.
Press BACK.
symbol means that OPEN LID AFTER
RUN is active.
Start delay - of time
Start centrifuging since
preselected delay is reached.
START DELAY/OF TIME
Via
keys mark START DELAY.
Press SET.
Via kays mark OF TIME.
Press SET -
appears.
Via
keys mark field 0 : 0 0 : 05 (for
example).
Press SET.
Start delay can be set from
0 : 0 0 : 0 1 to 9 : 5 9 : 5 9.
Confirm by pressing SET.
Press BACK.
When START DELAY-OF TIME function is
activated,
symbol is visible on the screen.
It is possible to exit START DELAY –of time mode at any time by pressing STOP key.