Service and Maintenance

5.1

Service of centrifuge
Ther is no routine maintenance or service required on the centrifuge other
than occasional cleaning.

5.2

General cleaning of the centrifuge
Always keep the centrifuge housing, rotor, and rotor accessories clean
by wiping down with a damp cloth.  Do not use harsh cleaners or
detergants on the lid or housing surfaces as it may affect the plastic.
After cleaning, ensure that all parts are dried thoroughly by hand or in a
warm-air-cabinet (max. temp. +50°C).

5.3

Disinfection
The centrifuge should be disinfected periodically and before shipping the
unit for service.  The centrifuge rotor area and rotor shields should be
wiped with alcohol or other disinfecting agents.  Do not allow the alcohol
or disinfectant to sit on the units lid or housing, as it may affect or
discolor the plastic.

5.4

Cleaning the centrifuge after breakage of glass tubes or bottles
With high g-values, there is a possibility that tube breakage can occur.
Should this happen the centrifuge rotor must be thoroughly cleaned
and all broken particles removed.  If this is not done, damage could
result to the centrifuge.Be sure to remove the rotor shields and clean
thoroughly.  If glass or other material has gotten into the centrifuge
enclosure, the enclosure must be removed for cleaning.

5.5

Replacing fuses
Check the fuse when it is recommended in the "Troubleshooting Guide".
The fuse holder is located underneath the unit.  Unplug the unit.  Unscrew
the fuse holder and remove the fuse.
Replace a blown fuse with a fuse of equivalant rating and type.

Содержание C0100-A

Страница 1: ...INSTRUCTION MANUAL FOR THE Z100 A CLINICAL CENTRIFUGE C0100 A January 2005...

Страница 2: ...3 2 4 CentrifugeInstallation 4 3 Operation 4 3 1 Closing the lid 4 3 2 Lidrelease 4 3 3 Lid lock 4 3 4 Startingthecentrifuge 4 4 Proper use of the Rotor 4 1 Overloading the rotor 5 4 2 Balancing the r...

Страница 3: ...without the prior permission of the manufacturer Exception tubes and bottles made of glass or plastic normally available in the laboratory and designed for centrifugation The centrifuge is to be used...

Страница 4: ...blocks until it has been established that your unit is working properly The accessories supplied with the centrifuge should be kept with the instruction manual near the centrifuge s place of installa...

Страница 5: ...ected in order to start the centrifuge The speed rpm can be selected between 1 000 and 4 000 rpm by turning the knob to the 1 2 3 and 4 speed marks To select a speed turn the speed knob to the 2 posis...

Страница 6: ...may operate a 6 place rotor with 2 3 or 4 loaded tubes but it is important that the occupied tube shields are symetrically arranged see sketch A 4 2 When loading the rotor you should proceed accordin...

Страница 7: ...cohol or other disinfecting agents Do not allow the alcohol or disinfectant to sit on the units lid or housing as it may affect or discolor the plastic 5 4 Cleaning the centrifuge after breakage of gl...

Страница 8: ...Check and set the time or speed Possible reason Blown fuse Solution Check fuse and replace if necessary see 5 5 Should you have any questions about the centrifuge and its accessories or require servic...

Страница 9: ...alues See section 1 3 for the correct radius of your rotor Note the use of tube adapters will change thisradius To determine the relative centrifugal force RCF value Thisvalueisthemultipliertothegravi...

Отзывы: