 - This manual is intended for use with the Z100-A

(C0100-A) Centrifuge.

1.1

Description
The Z100-A is a small bench top centrifuge designed for
sedimentation of various samples. The Z100-A has a
built in angle rotor capable of running up to to 4,000 rpm (1,900 x g.)
This product is for research only.

1.2

Safety precautions to be observed before operating the centrifuge

Do not operate if:

- The centrifuge has not been installed correctly.
- The centrifuge is partly dismantled.
- Service has been attempted by non-authorized or non-qualified

   personnel.
- Rotors and accessories are not part of the standard range of accessories
  or have been obtained without the prior permission of the

  manufacturer.
  Exception: tubes and bottles made of glass or plastic normally available
  in the laboratory, and designed for centrifugation.
- The centrifuge is to be used in explosive atmospheres.
- You intend to centrifuge combustible and explosive materials.
- You intend to centrifuge materials which are chemically reactive.

    If the equipment is used in a manner not specified by the manufacturer, the
    protection provided by the equipment may be impaired.

1.3 Technical Data

Model No.

C0100-A

Dimensions:
Width

295 mm (11.6 inch)

Depth

348 mm (13.7 inch)

Height

101 mm (10.1 inch)

Weight

5.5 kg / 12 lbs

Max. speed

4,000 rpm

Max. volume

90 ml (6 x 15ml)

Max. RCF

1,900 x g

Maximum Admiss. density

1.2 kg/dm

3

Rotor radius

10.62 cm

Timer

0 to 30 minutes or continuos

Electrical ratings

120 V/50-60 Hz, 1 Amp

The manufacturer reserves the right to make technical modifications

!

Содержание C0100-A

Страница 1: ...INSTRUCTION MANUAL FOR THE Z100 A CLINICAL CENTRIFUGE C0100 A January 2005...

Страница 2: ...3 2 4 CentrifugeInstallation 4 3 Operation 4 3 1 Closing the lid 4 3 2 Lidrelease 4 3 3 Lid lock 4 3 4 Startingthecentrifuge 4 4 Proper use of the Rotor 4 1 Overloading the rotor 5 4 2 Balancing the r...

Страница 3: ...without the prior permission of the manufacturer Exception tubes and bottles made of glass or plastic normally available in the laboratory and designed for centrifugation The centrifuge is to be used...

Страница 4: ...blocks until it has been established that your unit is working properly The accessories supplied with the centrifuge should be kept with the instruction manual near the centrifuge s place of installa...

Страница 5: ...ected in order to start the centrifuge The speed rpm can be selected between 1 000 and 4 000 rpm by turning the knob to the 1 2 3 and 4 speed marks To select a speed turn the speed knob to the 2 posis...

Страница 6: ...may operate a 6 place rotor with 2 3 or 4 loaded tubes but it is important that the occupied tube shields are symetrically arranged see sketch A 4 2 When loading the rotor you should proceed accordin...

Страница 7: ...cohol or other disinfecting agents Do not allow the alcohol or disinfectant to sit on the units lid or housing as it may affect or discolor the plastic 5 4 Cleaning the centrifuge after breakage of gl...

Страница 8: ...Check and set the time or speed Possible reason Blown fuse Solution Check fuse and replace if necessary see 5 5 Should you have any questions about the centrifuge and its accessories or require servic...

Страница 9: ...alues See section 1 3 for the correct radius of your rotor Note the use of tube adapters will change thisradius To determine the relative centrifugal force RCF value Thisvalueisthemultipliertothegravi...

Отзывы: