5.1.3. Connection with a three-sensor cable connector
Figure 11. Connection with a three-sensor cable connector
5.2. Explanations of connection diagrams
1
Supply voltage
11–17 V DC
(N.B. The device does not include a mains switch, see Section
General safety
instructions
)
-
= negative supply voltage
+
= positive supply voltage
Max. supply cable 2.5 mm
2
(-, +) AWG 13.
2
Sensor connectors
9
= sensor 1, connection 1
10 = sensor 1, connection 2
11
= sensor 2, connection 1
12 = sensor 2, connection 2
13 = sensor 3, connection 1
14 = sensor 3, connection 2
The sensor connection is a polarity-free digital bus, which means that it does not
matter which sensor terminal the wires are connected to.