For the Pt100 inputs,SMART probes can be used which incorporate dedicated
‘correction values‘ within the connector of each probe.When a SMART probe is
connected to input A or B or both inputs, the corrections are automatically applied
without the need for user intervention.
A common application for the Tempmaster PRO is the comparison calibration of
“working” sensors against a calibrated reference standard or semi-standard sensor. For
example, the sensor under test in channel B is compared against the reference sensor
(with programmed corrections) in channel A.
When applied, the null function corrects the differential temperature readout between
two sensors to zero. For example, the apparent temperature difference between two
sensors known to be at the same actual temperature can be corrected to zero prior
to, for example, heat exchanger experiments.
Both inputs can be scanned and values logged according to parameters set up by the
user in the PC software;the instrument incorporates a real-time clock and on-board
memory.
The adjustable contrast OLED display screen provides data readout, user prompts and
mode annunciation. This very important feature ensures a high degree of user
friendliness and confidence. User prompts indicate which buttons to press in the
appropriate sequence when selecting parameters and functions and when setting
calibration values. Mode annunciation indicates which mode of operation is currently
selected.
A flash drive USB port is provided to allow data to be stored and/or exported.
Firmware update are also facilitated via this port.
PC software running in WINDOWS is provided as standard; it allows programming of
custom calibration, remote control & measure and logging functions.
4