Chapter 6: Maintaining the Nano Enclosure 

 

Product Service 1-800-522-7658 

37 

 

Low Air Alarm 

Set Point 

(user defined) 

Enclosure 

Operating 

Inflow Speed 

Low Calibration 

Set Point  

(user defined) 

High Calibration 

Set Point  

(user defined) 

40 – 50 fpm 

60 fpm 

40 – 60 fpm 

100 – 120 fpm 

60 – 70 fpm 

80 fpm 

50 – 90 fpm 

100 – 150 fpm 

80 - 90 fpm 

100 fpm 

50 – 110 fpm 

100 – 170 fpm 

 
NOTE:   ENTER button stores information 
 

  +/- Buttons allow for scrolling. 

1.

 

Push the ENTER button on the face of the alarm until the Set up Menu is 
displayed. 

2.

 

Scroll to SET UP and hit ENTER. 

3.

 

The PASSWORD MENU displays (The Password is 0000).  Press the 
ENTER button repeatedly until the CAL CONFIG MENU is displayed. 

4.

 

In the CAL CONFIG MENU set for the following: 

 

CALIBRATION CONFIG MENU 

SETTINGS 

DISPLAY UNITS 

FPM 

LOW AIR ALARM 

SEE CHART ABOVE – LOW AIR ALARM 

LOW AIR CUTOFF 

OFF 

LOW AIR CUTOFF 

SEE CHART ABOVE – LOW AIR ALARM 

WARNING AIR ALARM 

SEE CHART ABOVE – LOW CALIBRATION SET 
POINT (Set 3-5 FPM above low air alarm) 

WARNING AIR RESET 

3 FPM 

HIGH AIR ALARM 

OFF 

HIGH AIR ALARM 

OFF 

LOWER AIR SAMPLE FLUCTUATIONS 

3% 

HIGHER AIR SAMPLE FLUCTUATIONS  3% 
LOWER/HIGHER AIR SAMPLE 
DIFFERENCE 

10 FPM 

WARN TO ALARM AIR TIME 

10 SECONDS 

ALARM TO WARN AIR TIME 

3 SECONDS 

SHOW AIR FLOW 

ON 

SHOW TIME LINE OFF = DISPLAYS 
BAR GRAPH 

OFF 

AUDIBLE ALARM 

ENABLED 

SENSOR DIFFERENCE 

2% 

 

5.

 

To complete the CAL CONFIG, be sure to enter “DONE”.  You are 
returned to the Main Menu. 

 

 

Содержание XPert 38872 Series

Страница 1: ...XPert Nano Enclosures Models 38872 Series 38873 Series 38874 Series 38875 Series 38876 Series 38885 Series 38886 Series...

Страница 2: ...rs of 7 30 a m and 5 30 p m Central Standard Time Part 3881800 Rev J ECO L352 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasona...

Страница 3: ...ctrical Supply Source to the Nano Enclosure 14 Set the Face Velocity with the Speed Control Adjustment 14 CHAPTER 4 HIGH PERFORMANCE FEATURES AND SAFETY PRECAUTIONS 16 High Performance Features 16 Saf...

Страница 4: ...er Replacement 41 Speed Control Replacement 42 Maintenance on the Optional Ionizer 43 Trace Odor Carbon Filter Installation 44 CHAPTER 7 ACCESSORIZING THE NANO ENCLOSURE 45 Guardian Airflow Monitors 4...

Страница 5: ...ake full advantage of them please acquaint yourself with this manual and keep it handy for future reference The XPert Nano Enclosures include a true bag in bag out ULPA filter to properly protect pers...

Страница 6: ...t in injury to the user or permanent damage to your Clean Bench Important information is presented in capitalized type in paragraphs that are preceded by the pointer icon It is imperative that the inf...

Страница 7: ...The exhaust and blower requirements The electrical power requirements The space requirements Refer to Appendix B Dimensions for complete enclosure dimensions Refer to Appendix C Specifications for co...

Страница 8: ...ates Do not place the balance and enclosure on a bench that would receive direct sunlight Humidity and Static Electricity Requirements Electrostatics can be troublesome in a balance enclosure It is im...

Страница 9: ...ms per square The table below lists the surface resistivity of various classes of material Surface Resistivity Table Material Surface Resistivity Example Conductive 0 105 per square Skin Metals Static...

Страница 10: ...CFM XPert Station Initial Static Pressure Loss with HEPA filter in w g XPert System Noise Pressure db A Xpert System Max Equivalent Resistance Ft Xpert System Max External Static Pressure 2 9 4 Sash 6...

Страница 11: ...onnecting the filtered enclosure or other equipment such as a balance for weighing operations The enclosures include iris pass thru to allow electrical cords through the back of the enclosure without...

Страница 12: ...th the speed control adjustment Arrange validation for the enclosure Depending upon which model you are installing you may need common mechanical and electrical installation tools in addition to wrenc...

Страница 13: ...OU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGES Do not discard the packing material until you have checked all of t...

Страница 14: ...d into its final location For stands with leveling feet use a carpenter level to adjust in both planes as shown in figure below 4 XPert Nano Enclosures have an integral stainless steel work surface wi...

Страница 15: ...as shown in Figure 3 3 For the 2 3 and 4 sizes the exhaust connection ships uninstalled on the top and needs to be fastened with the screws provided See Chapter 7 for ordering Canopy Exhaust Connecti...

Страница 16: ...Chapter 3 Getting Started Product Service 1 800 522 7658 12 Figure 3 2 FOR 6 NOMINAL EXHAUST DUCT CONNECTION 6 06 ID ON THE 2 3 AND 4...

Страница 17: ...Chapter 3 Getting Started Product Service 1 800 522 7658 13 Figure 3 3 10 NOMINAL ON THE 5 AND 6 10 06 ID...

Страница 18: ...elocity measurements are made using an anemometer An electric anemometer can be obtained from your laboratory supply dealer To determine the actual face velocity at the sash opening airflow velocity r...

Страница 19: ...hapter 3 Getting Started Product Service 1 800 522 7658 15 NOTE Face velocity profiles and smoke testing should be performed frequently per your organization s quality system to ensure safe performanc...

Страница 20: ...aerodynamic designs of the upper sash foil side air foils lower air foil upper dilution air supply and rear perforated baffle all work in concert to produce horizontal airflow patterns that significan...

Страница 21: ...Chapter 4 Performance Features and Safety Precautions Product Service 1 800 522 7658 17 Not Shown 9 10 11 12 17 18 19 20 21 23 24 Figure 4 1 22 14 8 13 3 2 15 1 5 6 7 4...

Страница 22: ...air foil air continues to pass under the air foil and through the Clean Sweep openings See Figures 4 1 and 4 3 The air foil is easily removable for cleaning Do not operate the enclosure without the a...

Страница 23: ...for loading and cleaning The sash must be down for normal operation The 5 and 6 units utilize a gas spring assist in place of the latch 7 Utility Ports allow electrical cords and data cords to pass th...

Страница 24: ...exhaust transition adapter designed for 6 diameter nominal duct on 2 3 and 4 sizes The 5 and 6 sizes use 10 nominal duct 21 Optional Ionizer not shown when activated floods the work area with positiv...

Страница 25: ...t the concentration of flammable or explosive gases or vapors Use of an open flame should be avoided in the enclosure as they may disrupt the airflow patterns in the cabinet Gases under high pressure...

Страница 26: ...enclosure 20 Proper operation of the enclosure depends largely upon the enclosure location and the operator s work habits The enclosure should be located away from traffic patterns doors fans ventila...

Страница 27: ...y raise the sash for loading and cleaning Check the baffle air slots for obstructions Allow the enclosure to operate unobstructed for 1 minute Wear a long sleeved lab coat and rubber gloves Use protec...

Страница 28: ...rticulate matter hazardous powders and biohazardous materials Because room air is drawn through the enclosure during operation this enclosure should not be used for operations requiring sample protect...

Страница 29: ...ntrol switch allows the blower to be operated independently of the ionizer Push the top of the switch to activate the blower only push the bottom to energize the ionizer along with the blower The gree...

Страница 30: ...otorized impeller and stainless steel interior should not be exposed to acids Where applications require the use of acids a separate ventilated enclosure or vented fume hood should be used with a remo...

Страница 31: ...ing the inflow face velocity Operating and calibrating the airflow monitors Initial certification Re certification Fluorescent light replacement Motorized impeller replacement Speed control replacemen...

Страница 32: ...on technician See Certification and Recertification in this chapter All monthly activities Decontamination When used in conjunction with biohazards the XPert Nano Enclosure should be decontaminated wi...

Страница 33: ...e Figures 4 1 and 4 2 4 Place both arms in the rear gloves of the bag and carefully draw the HEPA filter out of the enclosure Rest the HEPA filter on the floor table or cart Remove arms from bag glove...

Страница 34: ...n the bag Reinstall the front panel Figure 6 0 Figure 6 1 Bag In Bag Out Filter Changing Procedure Diagram Filter Leak Test Diagram Old Filter New Filter Old ULPA Filter Sealed in Old Bag Filter Gaske...

Страница 35: ...g during operation Air Techniques Inc Model TDA 4A or equal One nozzle at 10 psig is 67 5 cfm x 100ug l Vol of air For the 2 at 90 fpm or 130 cfm one nozzle 10 psig is 6 750 130 cfm 52 ug l For the 3...

Страница 36: ...ore than 1 inch from the surface of the filter media Scan the entire periphery of the filter and the gasket between the filter frame and the enclosure frame Scanning shall be done at a traverse rate o...

Страница 37: ...y time the airflow is safe good one can press the SILENCE ALARM test button and the audio alarm and the red LED will turn on as long as this button is held down The circuit board has also a two pin co...

Страница 38: ...ty Once the alarm condition is met simply increase the speed control outlined in Chapter 6 or replace the ULPA filter if the speed control is maximized 8 The table below lists typical alarm conditions...

Страница 39: ...Chapter 6 Maintaining the Nano Enclosure Product Service 1 800 522 7658 35 Figure 6 2 Guardian Airflow Monitor LED with Airflow Switch...

Страница 40: ...or external alarm and provides up to three output relays that can be configured For complete detailed information please refer to the separate Labconco 1000 Alarm User s Manual provided with the encl...

Страница 41: ...ENTER button repeatedly until the CAL CONFIG MENU is displayed 4 In the CAL CONFIG MENU set for the following CALIBRATION CONFIG MENU SETTINGS DISPLAY UNITS FPM LOW AIR ALARM SEE CHART ABOVE LOW AIR...

Страница 42: ...p out of the way wait approximately 30 seconds for air currents to settle and hit ENTER The airflow alarm will sample the airflow 5 Adjust the speed of the fan until the high airflow velocity is reach...

Страница 43: ...Chapter 6 Maintaining the Nano Enclosure Product Service 1 800 522 7658 39 Figure 6 3 Guardian Digital 1000 Airflow Monitor...

Страница 44: ...ghting Intensity Test when appropriate Noise Level Test when appropriate Vibration Test when appropriate Fluorescent Light Replacement 1 Disconnect the power 2 Locate the small light reflector located...

Страница 45: ...ventilateur tant qu il est en marche 1 Wear appropriate personal protective equipment to decontaminate the filtered enclosure and then unplug from the electrical outlet 2 Remove filter per the ULPA fi...

Страница 46: ...lding the speed control using a Phillips screwdriver Refer to Appendix A for Replacement Parts Diagram 3 Disconnect all wires leading to the speed control Connect wires on new speed control in the sam...

Страница 47: ...lectronic grade isopropyl alcohol and 50 de ionized water Emitter Point Replacement If the ionizer fails to neutralize charges in the airstream the only portions of the ionizer that can be serviced ar...

Страница 48: ...ed 1 Remove the screws 2 located on each side of the upper diffuser 2 Lift and remove the upper diffuser 3 Place the Trace Odor Carbon Filter on top of the HEPA filter with the gasket side DOWN Use ca...

Страница 49: ...Pert enclosures Description Order Part s Individual Parts Required Guardian Airflow Monitor 115V or 230V 1 ea 3811500 Airflow Monitor Printed Circuit Board 1 ea 3910700 Airflow Sensor w Connector 2 ea...

Страница 50: ...leaks See Figure 3 3 and Figure B 3 Order the following Part Description 3924400 2 Canopy Exhaust Connection 6 3924401 3 Canopy Exhaust Connection 6 3924402 4 Canopy Exhaust Connection 6 3924403 5 Can...

Страница 51: ...Blower 115V 60 Hz 4 5 A 720 950 760 1050 710 1060 620 1060 7068400 1 3 Remote Blower 115V 60 Hz 6 1 A 760 1050 710 1060 620 1060 430 1060 460 1150 500 1220 7068500 1 3 Explosion Proof Blower 115V 60...

Страница 52: ...5 and 6 have two each Filter Classification and Size Catalog X Pounds 2 Organic 18 x 18 x 1 3937200 5 5 lbs activated 3 or 6 Organic 30 x 18 x 1 3937300 9 2 lbs Activated 2 required on 6 4 Organic 42...

Страница 53: ...Provides a second level of filtration with a redundant filter in series Contact Labconco as kit include entire support system and bag in bag out system Order ULPA filter separate Catalog For use with...

Страница 54: ...al Ref Appendix D Enclosure has improper face velocity Have enclosure certified and check exhaust system Check filters for loading Adjust enclosure speed control Enclosure should have an average face...

Страница 55: ...ion steps outlined in this manual in Chapter 6 or in the supplied Digital 1000 Monitor Manual Constant audible alarm Check airflow and calibration of airflow monitor See Chapter 6 Continuous alarm Che...

Страница 56: ...Chapter 8 Troubleshooting Product Service 1 800 522 7658 52 Service Operating Record Log Customer Name Model Number Date Installed Serial Number Unit Location Date Comments...

Страница 57: ...2 Lamp Fluorescent 3 Enclosure F17T8 6C 1 9721901 Lamp Fluorescent 4 Enclosure F25T8 6D 1 9721900 Lamp Fluorescent 5 F32T8 6E 1 9721903 Lamp Fluorescent 6 F40T8 6F 1 9721910 LED 5 Enclosure 48 nominal...

Страница 58: ...5V 14B 1 3885302 Circuit Board 100V Ionizers 14C 1 3885300 Circuit Board 115V 230V Ionizers 14D 1 1325501 Speed Control 230V 15A 1 3885700 ULPA Filter 2 18 x 18 x 3 32 w Gasket 15B 1 or 2 3885701 ULPA...

Страница 59: ...Appendix A Replacement Parts Product Service 1 800 522 7658 55 15 16 10 12 14 2 1 5 6 7 9 3...

Страница 60: ...Appendix A Replacement Parts Product Service 1 800 522 7658 56 17 4 17 17 8 7 10 15...

Страница 61: ...Product Service 1 800 522 7658 57 Appendix B Dimensions See the following dimensions and exhaust options for all the XPert Nano Enclosures Figure B 1 XPert Nano Enclosure 2 3 and 4...

Страница 62: ...Appendix B Dimensions Product Service 1 800 522 7658 58 Figure B 2 XPert Nano Enclosure with Duct Connection or Canopy Connection 2 3 and 4...

Страница 63: ...Appendix B Dimensions Product Service 1 800 522 7658 59 Figure B 3 5 and 6 x 35 64 Deep XPert Nano Enclosure...

Страница 64: ...Appendix B Dimensions Product Service 1 800 522 7658 60 Figure B 4 5 and 6 x 29 13 Deep XPert Nano Enclosure...

Страница 65: ...Appendix B Dimensions Product Service 1 800 522 7658 61 Figure B 5 5 and 6 x 35 64 Deep XPert Nano Enclosure...

Страница 66: ...Appendix B Dimensions Product Service 1 800 522 7658 62 Figure B 6 5 and 6 x 29 13 Deep XPert Nano Enclosure...

Страница 67: ...ve humidity 80 for temperatures up to 88 F 31 C decreasing linearly to 50 relative humidity at 104 F 40 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Transient over voltag...

Страница 68: ...4x18x3 31 2 18x30x3 31 Motor Horsepower HP 0 13 0 13 0 30 0 60 0 60 Capacitor MFD Volts 10MFD 115V 2 4MFD 230V 10MFD 115V 2 4MFD 230V 15MFD 115V 4MFD 230V 15MFD 115V 4MFD 230V 15MFD 115V 4MFD 230V Flu...

Отзывы: