Appendix C:
Specifications
Electrical Data
Model #
Requirements
3044xxx0x
115 VAC, 60 Hz, 16 Amps
3044xxx1x
100 VAC, 50/60 Hz, 16 Amps
3044xxx-20, 30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
3046xxx0x
115 VAC, 60 Hz, 16 Amps
3046xxx1x
100 VAC, 50/60 Hz, 16 Amps
3046xxx-20, 30, 40, 50, 60, 70, 80 230 VAC, 50/60 Hz, 8 Amps
Motor Specifications
Cabinet Model
Electrical Requirements
All Cabinets, all
Voltages
2- 1/3 H.P. Electronically Commutated Motor (ECM)
120-240 VAC – 50/60 Hz,
Full Torque – 28 Oz.-Ft (3.56 N-M)
5 Full Load Amps @115VAC
2.6 Full Load Amps @230VAC
Automatic Thermal Protection
Environmental Conditions
Indoor use only.
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F
(31°C), decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient overvoltages according to Installation Categories II
(Overvoltage Categories per IEC 1010). Temporary voltage spikes on
the AC input line that may be as high as 1500V for 115V models and
2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally
only non-conductive atmospheres are present). Occasionally,
however, a temporary conductivity caused by condensation must be
expected, in accordance with IEC 664.