Chapter 2: Prerequisites
Product Service 1-800-522-7658
5
Exhaust Requirements
THE OPEN ACCESS STATION DISCHARGES ITS FILTERED
EXHAUST AIR UNDERNEATH THE UNIT. IT SHOULD BE
PLACED AT LEAST 12 INCHES AWAY FROM ANY
VERTICAL OBSTRUCTION TO OPERATE PROPERLY.
Electrical Requirements
The different PuriCare models have the following electrical requirements:
Model #
Requirements
3820000 & -01
115 VAC, 60 Hz, 12 Amps
3820010 & -11
100 VAC, 50/60 Hz, 12 Amps
3820020 & -21
230 VAC, 50/60 Hz, 6 Amps (US)
3820023 & -33
230 VAC, 50/60 Hz, 6 Amps (UK)
3820024 & -34
230 VAC, 50/60 Hz, 6 Amps (Schuko)
3820026 & -36
230 VAC, 50/60 Hz, 6 Amps (China/Australia)
3820028 & -38
230 VAC, 50/60 Hz, 6 Amps (India)
A dedicated outlet with a circuit breaker rated at 20 Amps for 115 volt models and
10 Amps for 230 volt models, should be located within 25 feet of the station for
115 volt models or within 10 feet for 230 volt models.
Figure 2-1
Table 2-1