Product Service 1-800-522-7658
48
A
PPENDIX
E
Q
UICK
C
HART
F
OR
T
HE
P
URI
C
ARE
O
PEN
A
CCESS
S
TATONS
Model
4FT 100 or 115V
4FT 230V
5FT 100 or 115V
5FT 230V
Sash Height (inches)
14
14
14
14
Downflow (FPM)
50-60
50-55
50-60
50-55
Approximate Settings for new Filters
Mag Gauge Reading
0.3-0.5
0.3-0.5
0.3-0.5
0.3-0.5
Supply Filter Number of Laskin Nozzles Needed
1
1
1
1
Exhaust Volume (CFM)
509-528
434-452
639-664
537-562
Exhaust Filter Number of Laskin Nozzles Needed
1
1
2
2
Supply HEPA Filter Dims. (in.)
42x18x3
42x18x3
48x18x3
48x18x3
Exhaust HEPA Filter Dims (in.)
36x18x6
36x18x6
48x18x6
48x18x6
Fluorescent Light(s)
2 each
F25T8/TL741
2 each
F25T8/TL741
2 each
F32T8/TL741
2 each
F32T8/TL741