Chapter 4: Performance Features and Safety Precautions
Product Service 1-800-522-7658
18
HEPA filters only remove particulate matter. Operations generating volatile toxic
chemicals are prohibited. The media of the filter is fragile and should not be
touched. Avoid puncturing the HEPA filter during installation or normal
operation. If you suspect that a HEPA filter has been damaged, DO NOT use the
cabinet; contact Labconco at 800-821-5525 or 816-333-8811 for further
information.
The HEPA filters in the cabinet will gradually accumulate airborne particulate
matter from the items placed in the cabinet. The rate of accumulation will depend
upon the cleanliness of the air, operating time and the nature of items being dried
in the cabinet.
Do not operate the cabinet without all of the appropriate filters in place. If an odor
is detected coming out of the granular activated carbon, it needs to be replaced.
The weight limit for the fold down shelves is 15 lbs (6.8 kg), and the floor of the
cabinet and lower storage area is 20 lbs (9.1 kg).
When surface disinfecting the cabinet:
Always wear appropriate personnel protective equipment.
NEVER use high pressure water to clean the liner; this practice could
generate biohazardous aerosols.
NEVER spray disinfectants into the blower intake at the top of the liner;
this may damage the blower.
Avoid splashing the disinfecting solution on skin or clothing.
Ensure adequate room ventilation.
Carefully follow the disinfectant’s safety instructions.
Always dispose of disinfecting solutions in accordance with local and
national laws.
DO NOT allow bleach or disinfectants with high concentrations of free
chlorine to contact the stainless steel components of the cabinet for a
long period of time. Free chlorine will corrode stainless steel after
extended contact.
When handling used filters or prefilters, always wear appropriate personnel
protective equipment.