Chapter 3: Getting Started
Product Service 1-800-522-7658
9
Drain Options - UV Lamp Models
(Model # 34000-00 thru -04,
34040-00 thru -04 & 34050-00 thru -04 only)
NOTE: These models can either drain to a bucket, already located in the
lower storage area, or the bucket can be removed, and an optional Drain Kit
(#3928200) can be installed. The optional drain kit allows the cabinet to drain
to a nearby floor drain.
On units equipped with the drain bucket:
NOTE: The drain bucket had a working capacity of approximately 1.5
gallons. Periodically monitor the liquid level in the bucket during
decontamination or washdown operations.
1.
Access the bucket by opening the bottom storage area door on the front of
the cabinet.
2.
Before using, ensure the bucket is in its correct position under the drain.
To install the Drain Kit (#3928200):
NOTE: The drain line should be routed to a floor drain with an air gap
only. NEVER directly connect the drain line to a sewer or drain system, or
use a stand pipe, to prevent backflow into the cabinet.
1.
Remove the bucket from the lower storage area and save for possible reuse.
2.
Connect the drain hose assembly to the tail pipe of the drain (on the bottom
of the drain assembly). Push the drain assembly up onto the stand pipe until
it is fully seated.
3.
Hand-tighten the compression ring of the drain assembly to the tail pipe
until it is secure.
4.
Route the drain hose through the appropriate hole in the bottom rear panel
of the cabinet.
5.
Route the drain hose to the floor drain and secure the end of the hose to
prevent it from moving.
6.
Pour water into the cabinet drain, and inspect the entire drain system for
leaks. Tighten any leaking fittings.