Chapter 7: Accessorizing Your Glove Box
Product Service 1-800-522-7658
39
14.
Filtration Components for Moisture, Oxygen and Fume Removal
The following components can be used to remove organics, acids,
ammonia, oxygen, moisture and other gases as indicated. They are
connected to the two auxiliary valves on the main glove box chamber
for recirculation of main chamber atmosphere and contaminant
removal of air (see Figures 3-9 and 3-10).
Part #
Descriptions
5248200
Vacuum Pump for Circulation, 115V, 60 Hz w/switch
5248201
Vacuum Pump for Circulation, 230V, 50 Hz International
7815301
Canister with Stand for Traps
7815200
Moisture & Solvent Trap Insert; to provide less than 50 ppm
moisture (molecular sieve)
7814800
Acid Vapor Trap Insert
7815000 Radioisotope
Trap
Insert
7995600
Ammonia Trap Insert
5244100 Flowmeter,
Visual
5244200 Oxygen
and
Moisture Removal Column; to provide less than
5 ppm oxygen and less than 50 ppm moisture (refillable)
5244300
Oxygen and Moisture Removal Column Support Stand
15.
Drying Train Tubing Kit
Includes the (FEP) tubing and connections required for connecting
typical drying train components such as moisture, oxygen, and other
gases. Fluorinated ethylene-propylene (FEP) tubing has very low
moisture absorption.
Part #
Descriptions
5242500
Drying Train Tubing Kit; moisture, solvents and
radioisotopes
5242501
Drying Train Tubing Kit; oxygen and moisture only
See Figures 3-9 and 3-10
16.
Gas and Vacuum 0.2 Micron Filter (Part #5240100)
Used for removal of particulates in a drying train, vacuum pump
exhaust or inert gas line. 0.25"-0.38" ID tube connection with
additional fittings for attachment to 0.50" ID hose.
17.
Oxygen Monitors
Oxygen monitors allow the user to continuously monitor oxygen
levels inside the glove box. All monitors are placed inside the main
chamber.
Never operate a transfer chamber vacuum purge on
these sensitive oxygen monitors as damage can occur.
Part #
Descriptions
5244500
0-10,000 ppm Oxygen Monitor (with three ranges of 0-100 ppm, 100-
1,000 ppm, and 1,000-10,000 ppm) 100V/115V – 50/60 Hz
5244501
0-10,000 ppm Oxygen Monitor (with three ranges of 0-100 ppm, 100-
1,000 ppm, and 1,000-10,000 ppm) 230V – 50/60 Hz – International
5244400
0.1-100% Oxygen Monitor (1,000 ppm lowest) 100V/115V/230V –
50/60 Hz