Chapter 6: Troubleshooting
Product Service 1-800-522-7658
40
Original instructions
4.
The assembly is now ready for repairs. Whenever possible, do not break the
fluid system.
5.
Reverse the removal procedures for installation of the shelf and stoppering
system. Make sure the fluid inlet and outlet tubes are routed on top of the
assembly.
Filling the Fluid System
1.
Fluid Circulation Valve
2.
Fluid Reservoir and Cap
3.
Electrical Junction Block
4.
Flexible Tubing
1.
Make sure the Power Switch is in the OFF position.
2.
Remove the left cover.
3.
Remove the insulation and both caps from the fluid circulation valve,
which is located directly behind the fluid reservoir.
4.
Attach a flexible tube to the exposed port (service port) of the fluid
circulation valve. The port is 3/8" x 45° male flare.
5.
Remove the fluid reservoir cap and place the opposite end of the tubing
into the fluid reservoir.
6.
Fill the reservoir slowly allowing the fluid to drain through the system.
When the system appears full, prepare to start the circulation pump.
2
1
4
3