Installing a Secondary Chemical Trap
An accessory secondary chemical trap is available to minimize the exhausting of
solvents into the atmosphere. Attach the bracket to the right hand side of the
housing using the screws provided. Attach the hose from the outlet on the
vacuum pump to the “out” connector of the canister housing. Unscrew the clear
blow of the canister housing from the head. Remove both the upper and lower
caps from the filter cartridge and insert the small end of the cartridge into the hole
in the center of the head. Reinstall the clear bowl.