![Labconco CentriVap 78110-00 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/labconco/centrivap-78110-00/centrivap-78110-00_user-manual_3285116028.webp)
24
Product Service Domestic 1-800-522-7658, International 816-333-8811
Chapter 4: Using Your CentriVap
Change heat set point:
To change the heat set point, press the set point
“SELECT” button until arrows point to the temperature
set point which can be changed from -- (OFF) up to 99°
by pressing the increase or decrease buttons. The set
point is 100°C if the display shows “HI.” The last
entered set point is stored in memory.
Change run time or heater time set point:
To change the time set point, press the set point
“SELECT” button until arrows point to the run time set
point which can be changed from 1 to 999 minutes by
pressing the increase or decrease buttons. If it is
desired to have the CentriVap run continuously without
alarming at the end of a time period, press the increase
button until the display says “ON.” The last entered set
point is stored in memory.
If during a run the “STOP” button is pressed, the timers
remember the time at which stop occurred. If “RUN” is
then pressed, the timers continue to count down from
the time at which they were stopped. If you are running
program 1, program 2 or program 3, pressing “PROG
1”, “PROG 2” or “PROG 3” resets the timers to the
original set point time and the CentriVap starts a new
run.
To reset the timers to the original set point time when
running programs other than program 1, program 2 or
program 3, press the start button and hold it for five
seconds. The display will indicate that the timer is
reset.
Operating The CentriVap
Never lift the lid if the rotor is moving.
Never attempt to reach into the
chamber if the rotor is moving.
Never attempt to defeat the lid latch
mechanism.
!!!!
Содержание CentriVap 78110-00
Страница 43: ...41 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A CentriVap Components...
Страница 45: ...43 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A CentriVap Components...
Страница 47: ...45 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A CentriVap Components...
Страница 63: ...63 Product Service Domestic 1 800 522 7658 International 816 333 8811...