14
Product Service Domestic 1-800-522-7658, International 816-333-8811
Chapter 3: Getting Started
wall surfaces. Restriction of the airflow through the
cabinet during operation could adversely affect
performance.
After positioning the components, it is necessary to join
the system together using the hoses provided. Attach
one hose to the .50 diameter tube that extends out the
back of the Concentrator. Attach the other end of this
hose to one of the barb fittings on the Cold Trap Cover
Assembly. Clamp securely with the clamps supplied.
Attach another hose to the remaining barb fitting on the
Cold Trap Cover Assembly and clamp securely. If the
accessory Secondary Trap is not used, attach the other
end of the hose to the inlet port on the vacuum pump.
If the accessory Secondary Trap is used, install the
Secondary Trap as explained in
Chapter 6: Modifying
Your CentriVap
. Then attach the hose from the Cold
Trap Cover Assembly to the “out” connector on the
Secondary Trap
Concentrator
Cold Trap
Vacuum Pump