Refer to the following if your CentriVap fails to operate
properly. If the suggested corrective actions do not
solve your problem, contact Labconco for additional
assistance. The following failure codes may appear on
the display when problems are sensed by the internal
self check routine.
DISPLAY
ERROR CODE
CAUSE CORRECTIVE ACTION
HEAT SENSOR
Sensor failure
Replace sensor assembly.
Connection failure
Repair connection.
CLOSE LID
Lid open
Close lid.
LATCH FAIL
Solenoid failure
Check component.
Switch or sensor
failure
Check connections.
MEM FAIL P1
Memory failure
Push program button #1
New memory IC chip
Push program button #1.
Bad memory IC chip
Call Labconco – Replace IC chip or control PCB.
MOTOR ERROR
Defective motor
Replace motor.
Hall effect sensor
failure
Replace sensor.
Wire failure
Replace wire.
Содержание CentriVap 74600-00
Страница 43: ...41 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A CentriVap Components...
Страница 45: ...43 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A CentriVap Components...
Страница 47: ...45 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A CentriVap Components...
Страница 63: ...63 Product Service Domestic 1 800 522 7658 International 816 333 8811...