WARNING
:
•
This centrifuge is not explosion-proof. Never use explosive or flammable samples.
•
Do not install the centrifuge in or near places where inflammable gases are generated or chemicals are
stored.
•
Do not place dangerous material within 30cm around the centrifuge.
•
Make sure to prepare necessary safety measures before using samples that are toxic, radioactive or
contaminated with pathogenic micro-organisms at your own responsibility.
•
If the instrument, rotor and/or accessories that has been contaminated by solutions with toxic, radioactive or
pathogenic materials, clean it according to the decontamination procedure that you are specified.
•
If you require services at site, please sterilise and decontaminate it in advance, and then notice the service
centre involved in the details of the particular materials.
•
Do not handle the power cord or turn on or off the POWER switch with wet hands to void electrical shocks.
•
For safety purposes, do not enter within 30cm around this centrifuge while it is in operation.
•
While the rotor is rotating, never forcedly release the door lock.
•
Unauthorized repairs, disassembly, and other services to the centrifuge except by our service centre are
strictly prohibited.
•
CAUTION:
•
This centrifuge must be located on one firm and level table.
•
Make sure the centrifuge is horizontal before running.
•
Make sure the angle between the door and cover is greater than 70 degrees when open the door.
•
Be careful not to put your fingers or hands between the door and cover when the door is off.
•
Do not move or relocate this centrifuge while it is running.
•
When there is fluid in the rotor chamber, please promptly dry with a dry cloth to avoid sample contamination.
•
Ensure to remove any objects and fragments of the tubes dropped inside the rotor chamber before running this
centrifuge.
•
Caution on rotors
o
Always check for corrosion and damages on the rotor surface before using it. Do not use the rotor if
such abnormality found.
o
Do not set the centrifuge speed override the allowable minimum speed of the rotor kits (rotor or the
other adapters). Make sure to run it below the allowable minimum speed.
o
Do not exceed the allowable imbalance.
o
Use the rotor and tubes within their actual capacities.
o
If the rotor is provided with a lid, ensure it is tightened before the operation.
•
If any abnormal condition occurs during operation, please stop it immediately and contact our service centre.
Notice the service the alarm code if displayed.