11.1.2 AVAILABLE ROTORS AND ADAPTORS 

The rotor can be used as follows: 

ROTOR 

TYPE 

ID CODE 

TUBE 

ADAPTER 

MAX 

SPEED (RPM) 

MAX RCF (

×

G) 

ALLOWABLE IMBALANCE(*) 

Imbalance 

Volume 

Imbalance 

400.

003.

055

 

01

 

1.5/2.0ml 

tube

 

 

15000

 

15100

 

8.5 

18620  

 

0.2ml 
PCR tube 

400.003.065

 

15000

 

15100

 

6.9  

15110  

 

0.5ml 
microtube 

400.003.066

 

15000

 

15100

 

7.6  

 

16638  

 

Table 11.1 Rotors and adapters

 

Notice 

 

The centrifuge rotor can separate sample which density lower than 2.0g/ml. if the separated samples density is over 

2.0g/ml, please calculate allowable speed depending on the following formula.  

 

Allow Speed (rpm)= Maximum speed

×

(2.0(g/ml)/Sample density (g/ml))

1/2

 

Autoclaving 

 

The rotor is made of high-strength aluminium alloy material, can be high-pressure sterilization: 121

1.0kg/cm

2

20 minutes. 

Cleaning and sterilising tubes please refer to the table below 

PA: Polyallomer; PC: Polycarbonate; PP: Polypropylene | O: Applicable X: Inapplicable 

CONDITION 

MATERIAL 

PA 

PC 

PP 

CL
E

A

N

IN

G

 

Cleaning Fluids 

Acidic

pH5 or lower

 

Acidic

higher than pH5 

 

Alkaline

higher than pH9 

 

Alkaline

pH9 or lower

 

Neutral

pH7

 

Warm water(up to 70

Ultrasonic cleaning 

Neutral detergent

pH7

 

STER

IL

IZA
TI

O

N

 

Autoclaving 

115

0.7kg/cm

2

30minutes 

121

1.0kg/cm

2

20 minutes 

126

1.4kg/cm

2

15 minutes 

Boiling 

15 to 30 minutes 

Ultraviolet sterilization 

200-300nm 

Gas sterilization 

Ethylene oxide 

Formaldehyde 

Table 11.2 Cleaning and sterilizing conditions for tubes 

 CAUTION: The lid of the rotor is made of plastics, cannot be high-pressure sterilisation, only ordinary sterilisation can 

be used. 

Содержание 400.003.050

Страница 1: ...MINI CENTRIFUGE USER MANUAL sales labcoscientific com au 1800 052 226 ABN 57 622 896 593 labcoscientific com au PO Box 5816 Brendale QLD 4500 Version 1 1...

Страница 2: ...FICATIONS 5 2 DECLARATION OF CONFORMITY 5 3 REQUIRED OPERATIONAL CONDITIONS 6 4 INSTALLATION 6 5 STRUCTURE 7 6 OPERATION PANEL 8 7 ROTOR PREPARATION 9 8 OPERATION 10 9 MAINTENANCE 14 10 TROUBLESHOOTIN...

Страница 3: ...as been operated and maintained properly Warranty of the Rotor Please pay attention do not use the rotor once it has been corrosion or fatigue damage We do not guarantee this centrifuge and the rotor...

Страница 4: ...centrifuge in any manner not described in this User manual When in doubt or have any troubles with this centrifuge ASK FOR HELP The precautions described in this User manual are carefully developed i...

Страница 5: ...ices to the centrifuge except by our service centre are strictly prohibited CAUTION This centrifuge must be located on one firm and level table Make sure the centrifuge is horizontal before running Ma...

Страница 6: ...omplies with Canadian ICES 001 SPECIFICATIONS DESCRIPTION MAXIMUM SPEED 15000rpm increment 100rpm MAXIMUM RCF 15100 g increment 100 g MAXIMUM CAPACITY 2ml 12 TIMER 30 seconds 99 minutes HOLD continuou...

Страница 7: ...ge the rotor must be removed Location 1 Locate this centrifuge on a firm flat and level table ensure the four feet of this centrifuge stand on the table firmly Avoid installing on a slippery table boa...

Страница 8: ...utlet 2 A cover 10A outlet which has a good ground is required and meets with local safety requirements 5 STRUCTURE Figure 5 1 400 003 050 front view of centrifuge Door lock hook Rotor Operation panel...

Страница 9: ...8 Figure 5 2 D2012 Rear view of centrifuge 6 OPERATION PANEL Speed area Lock status Time area 150 x 100rpm 12 min 1 2 3 4 5 Figure 6 1 Operation Panel Air vents View Door Power inlet Power switch...

Страница 10: ...utton Press down the button to start running The centrifuge will stop running if the button is pressed during centrifugation 1 Speed area Lock status Time area 150 x 100rpm 12 min Figure 6 2 the main...

Страница 11: ...m 12 min Figure 8 1 the last running interface o Speed 15000rpm Running time 12 minutes o The door lock is released CAUTION Any abnormalities such as corrosion or scratches are found stop using the ro...

Страница 12: ...firmly Put the rotor lid on the rotor press the fastener down and make the fastener join with the rotor nut tightly Close the door and start running To release the rotor first pull the fastener up to...

Страница 13: ...m minimum counter clockwise rotate parameter button values circulate from large small minimum maximum 2 Set the time Press select button time value flash into time setting mode Rotate parameter button...

Страница 14: ...ol will flash into RCF parameter input status If no button is pressed after the speed parameter has flashed for 5 seconds the inputting mode will be shut down 1 Rotate parameter button to input a RCF...

Страница 15: ...can wipe the drive shaft with soft cloth and then apply a thin coat of silicon grease to it 4 Door Clean and sterilise the door using the same method as the step 1 5 Rotor To prevent corrosion remove...

Страница 16: ...e door opening Close the door immediately Close the door and then start to operate E 06 Set wrong speed The setting speed exceeds the allowable range Modify the speed value E 10 86 Read the service ma...

Страница 17: ...on the top right side of the unit Insert a door key into the hole and push forward to release the door lock and then open the door 11 INSTRUCTIONS OF ROTOR AND TUBE The rotor instructions Rotor struct...

Страница 18: ...can be high pressure sterilization 121 1 0kg cm2 20 minutes Cleaning and sterilising tubes please refer to the table below PA Polyallomer PC Polycarbonate PP Polypropylene O Applicable X Inapplicable...

Страница 19: ...ure before the tubes are removed 11 2 4 CONDITION AND LIFE EXPECTANCY OF TUBES The life expectancy of plastic tubes depends on the characteristics of samples speed of the rotor used and temperature ap...

Страница 20: ...19...

Отзывы: