11.1.2 AVAILABLE ROTORS AND ADAPTORS
The rotor can be used as follows:
ROTOR
TYPE
ID CODE
TUBE
ADAPTER
MAX
SPEED (RPM)
MAX RCF (
×
G)
ALLOWABLE IMBALANCE(*)
Imbalance
Volume
Imbalance
400.
003.
055
01
1.5/2.0ml
tube
15000
15100
8.5
18620
0.2ml
PCR tube
400.003.065
15000
15100
6.9
15110
0.5ml
microtube
400.003.066
15000
15100
7.6
16638
Table 11.1 Rotors and adapters
Notice
•
The centrifuge rotor can separate sample which density lower than 2.0g/ml. if the separated samples density is over
2.0g/ml, please calculate allowable speed depending on the following formula.
•
Allow Speed (rpm)= Maximum speed
×
(2.0(g/ml)/Sample density (g/ml))
1/2
Autoclaving
•
The rotor is made of high-strength aluminium alloy material, can be high-pressure sterilization: 121
℃
(
1.0kg/cm
2
)
20 minutes.
Cleaning and sterilising tubes please refer to the table below
PA: Polyallomer; PC: Polycarbonate; PP: Polypropylene | O: Applicable X: Inapplicable
CONDITION
MATERIAL
PA
PC
PP
CL
E
A
N
IN
G
Cleaning Fluids
Acidic
(
pH5 or lower
)
X
X
X
Acidic
(
higher than pH5
)
O
O
O
Alkaline
(
higher than pH9
)
O
X
O
Alkaline
(
pH9 or lower
)
O
O
O
Neutral
(
pH7
)
O
O
O
Warm water(up to 70
℃
)
O
O
O
Ultrasonic cleaning
Neutral detergent
(
pH7
)
O
O
O
STER
IL
IZA
TI
O
N
Autoclaving
115
℃
(
0.7kg/cm
2
)
30minutes
O
O
O
121
℃
(
1.0kg/cm
2
)
20 minutes
X
O
O
126
℃
(
1.4kg/cm
2
)
15 minutes
X
X
X
Boiling
15 to 30 minutes
O
O
O
Ultraviolet sterilization
200-300nm
X
X
X
Gas sterilization
Ethylene oxide
O
X
O
Formaldehyde
O
O
O
Table 11.2 Cleaning and sterilizing conditions for tubes
CAUTION: The lid of the rotor is made of plastics, cannot be high-pressure sterilisation, only ordinary sterilisation can
be used.