8.1.2. SET THE OPERATION PARAMETERS 

Press the 

 button to select required parameters. The parameter can be modified when the parameter is flashing. 

Clockwise rotate the parameter button

to increase parameter value; counter-clockwise rotate the parameter button 

 

to decrease parameter value. The minimum speed increment is 100 rpm, the minimum time increment is 1 second. 

1.

 

Set the speed 

 

Press the select button

, choose the speed unit rpm, showing the speed parameter. 

 

While select speed button, the speed symbol will flash into the speed parameter input status. 

 

The minimum speed value you can set 500rpm, the minimum increment is 100rpm. 

 

Clockwise rotate parameter button

to increase parameter value; counter-clockwise rotate parameter button

to decrease parameter value. 

 

You can quick set value by parameter button

, quickly rotate parameter button

 to speed up parameter 

changing. 

 

There is a circulating function to increase/decrease parameters. Clockwise rotate parameter button

, values 

circulate from small 

 large 

 maximum 

 minimum; counter-clockwise rotate parameter button

, values 

circulate from large 

 small 

 minimum 

 maximum. 

2.

 

Set the time 

 

Press select button

, time value flash into time setting mode. 

 

Rotate parameter button

 to set running time range from 30 seconds to 99 minutes. 

 

When time displays as HD, that showing long-running mode. 

 

8.1.3. START THE OPERATION 

1. Press running button

 to start running 

 

It will start timing once the rotor starts running; the screen displays remaining run time. 

2. Inquiry and modify the operation parameters 

 

Press select button

, return to the preparation interface and displays setting parameters. Press again the 

select button

, once the parameter symbol is flashing, rotate parameter button

 to modify parameters. 

Release button after 5 seconds, centrifuge will return to normal operation mode and run process according to the 
new parameters. 

 

After the time parameter has been modified, the operation time is not cleared and will continue to be 

accumulated. 

3. Alarm display 

 

If anything is wrong in processing, the centrifuge will brake to stop automatically, and display the error code on 

the time area. You can find out the malfunction causes by checking the table 10-1, and make correct actions 
accordingly. 

Содержание 400.003.050

Страница 1: ...MINI CENTRIFUGE USER MANUAL sales labcoscientific com au 1800 052 226 ABN 57 622 896 593 labcoscientific com au PO Box 5816 Brendale QLD 4500 Version 1 1...

Страница 2: ...FICATIONS 5 2 DECLARATION OF CONFORMITY 5 3 REQUIRED OPERATIONAL CONDITIONS 6 4 INSTALLATION 6 5 STRUCTURE 7 6 OPERATION PANEL 8 7 ROTOR PREPARATION 9 8 OPERATION 10 9 MAINTENANCE 14 10 TROUBLESHOOTIN...

Страница 3: ...as been operated and maintained properly Warranty of the Rotor Please pay attention do not use the rotor once it has been corrosion or fatigue damage We do not guarantee this centrifuge and the rotor...

Страница 4: ...centrifuge in any manner not described in this User manual When in doubt or have any troubles with this centrifuge ASK FOR HELP The precautions described in this User manual are carefully developed i...

Страница 5: ...ices to the centrifuge except by our service centre are strictly prohibited CAUTION This centrifuge must be located on one firm and level table Make sure the centrifuge is horizontal before running Ma...

Страница 6: ...omplies with Canadian ICES 001 SPECIFICATIONS DESCRIPTION MAXIMUM SPEED 15000rpm increment 100rpm MAXIMUM RCF 15100 g increment 100 g MAXIMUM CAPACITY 2ml 12 TIMER 30 seconds 99 minutes HOLD continuou...

Страница 7: ...ge the rotor must be removed Location 1 Locate this centrifuge on a firm flat and level table ensure the four feet of this centrifuge stand on the table firmly Avoid installing on a slippery table boa...

Страница 8: ...utlet 2 A cover 10A outlet which has a good ground is required and meets with local safety requirements 5 STRUCTURE Figure 5 1 400 003 050 front view of centrifuge Door lock hook Rotor Operation panel...

Страница 9: ...8 Figure 5 2 D2012 Rear view of centrifuge 6 OPERATION PANEL Speed area Lock status Time area 150 x 100rpm 12 min 1 2 3 4 5 Figure 6 1 Operation Panel Air vents View Door Power inlet Power switch...

Страница 10: ...utton Press down the button to start running The centrifuge will stop running if the button is pressed during centrifugation 1 Speed area Lock status Time area 150 x 100rpm 12 min Figure 6 2 the main...

Страница 11: ...m 12 min Figure 8 1 the last running interface o Speed 15000rpm Running time 12 minutes o The door lock is released CAUTION Any abnormalities such as corrosion or scratches are found stop using the ro...

Страница 12: ...firmly Put the rotor lid on the rotor press the fastener down and make the fastener join with the rotor nut tightly Close the door and start running To release the rotor first pull the fastener up to...

Страница 13: ...m minimum counter clockwise rotate parameter button values circulate from large small minimum maximum 2 Set the time Press select button time value flash into time setting mode Rotate parameter button...

Страница 14: ...ol will flash into RCF parameter input status If no button is pressed after the speed parameter has flashed for 5 seconds the inputting mode will be shut down 1 Rotate parameter button to input a RCF...

Страница 15: ...can wipe the drive shaft with soft cloth and then apply a thin coat of silicon grease to it 4 Door Clean and sterilise the door using the same method as the step 1 5 Rotor To prevent corrosion remove...

Страница 16: ...e door opening Close the door immediately Close the door and then start to operate E 06 Set wrong speed The setting speed exceeds the allowable range Modify the speed value E 10 86 Read the service ma...

Страница 17: ...on the top right side of the unit Insert a door key into the hole and push forward to release the door lock and then open the door 11 INSTRUCTIONS OF ROTOR AND TUBE The rotor instructions Rotor struct...

Страница 18: ...can be high pressure sterilization 121 1 0kg cm2 20 minutes Cleaning and sterilising tubes please refer to the table below PA Polyallomer PC Polycarbonate PP Polypropylene O Applicable X Inapplicable...

Страница 19: ...ure before the tubes are removed 11 2 4 CONDITION AND LIFE EXPECTANCY OF TUBES The life expectancy of plastic tubes depends on the characteristics of samples speed of the rotor used and temperature ap...