Page 25
② If you click
View → Status,
additional separate window appears below showing actual
temperature and set point window. Operating, Auto Tune, Program, Over Temp., Level,
and etc display on this window.
The following picture is a monitoring window after choosing Status and Parameter in
View menu.
User can monitor operating process through three divided interfaces. (Graph,
Parameter, Status)
☞ Graph
displays;
Actual temperature (Red line) and set point (Blue line) on the top of separate
interfaces.
☞
Status
says that;
-
Operating
represents the unit is Running.
If blue line is Hi, the unit is On(operating). If blue line is Low, the unit is Off.
-
Auto Tune
displays whether the unit performs Auto Tuning or not.
-
Program
displays whether the unit is in programmable mode or not.
-
Over Temp
. displays over heating condition of a unit.
-
Water Low
displays whether Water Level works or not.
In case water is at the low level, blue line is Hi position and low position under
normal condition.
-
Cooling
displays whether compressor works or not.