5. Unit operation
1) Basic operation
(1) Put in the sample in the oven and close the door. (Do not turn on the oven.)
(2) Connect vacuum pump and the oven to create vacuum. (Please refer to p8.)
(3) Switch on the power.
(4) Set the Over Temp. Protector Knob higher than operating temperature. (about 20%).
(5) Set temperature.
(6) Press Start / Stop Key to start operation.
2) Temperature setting method
(1) Press button.
Setting temperature value (SV) should blink. This
means you can vary set value.
(2) Press button to change value then
press button to fix the value.
(3) It reverts to the previous value without saving if you
don’t touch any button for 10 seconds.
(4) Press button again and when it is in SV set
state then following additional functions are
displayed by putting above key again.