![L3Harris NARDA PMM ER8000 Скачать руководство пользователя страница 69](http://html.mh-extra.com/html/l3harris/narda-pmm-er8000/narda-pmm-er8000_user-manual_665120069.webp)
Remote control
7-17
SSFW
n,freq
This setting command is used for creating a Frequency-Scan-Tab frequency by
frequency. It is made of two fields as follows:
•
n
Is the index of frequency being written and it should range from 0 to 99
•
freq
Is the frequency, expressed in Hz (exponential notation allowed), being
written
The procedure to make and use a Frequency-Scan-Tab on PMM ER8000 is the
following:
•
Send as many commands
SSFW
as frequency points of table are in
upwards order.
•
Send the command
SSFD
with step size set to zero
Here is an example how to make a Frequency-Scan-Tab on PMMER8000:
•
#SSFW 0, 150e3 * Set the first freq to 150kHz
•
# SSFW 1, 500e3 * Set the second freq to 500kHz
•
# SSFW 2, 5e6* Set the third freq to 5MHz
•
# SSFW 3, 6e6* Set the fourth freq to 6MHz
•
# SSFW 4, 30e6 * Set the fifth freq to 30MHz
Note that all the higher indexes in the Frequency-Scan-Tab are cleared. Therefore, a
Frequency-Scan-Table must be made in upwards order. All points, having a higher
index previously written, are cleared.
Note that only the present frequencies will be tuned and measured as all remaining
will be completely neglected. Hence, it is
responsibility of the user to make sure
the data sent are correct and coherent.
The reply is
SSFW
=OK
which acknowledges the command has been granted or
SSFW
=SERR
if the command has been ignored.
SSTP
This setting command stops abruptly a running sweep in Analyzer mode. The reply is
always
STP=OK,
which acknowledges the command has been granted
STAT b
This setting command sets MinAtt to the value indicated by (b) which range from 0dB
to the maximum available attenuation
and must be in step of 5dB. The reply is
TAT=OK
which acknowledges the command has been granted or
TAT=SERR
if the
command has been ignored.
Example:
#STAT 10*
Note: In Sweep mode the Attenuator is always forced to AUTO
SUPP
n
This setting command
#SUPP n*
outputs on the User Port bit by bit the argument n .
Each Pin has its own contribution according to its index as follows:
OUT0= 2^0
OUT1= 2^1
OUT2= 2^2
Thus, the argument sent is the sum of each weighted contribution
The reply is
UPP=OK
which acknowledges the command has been granted or
UPP=SERR
if the command has been ignored (IE argument >31).
Example
: SUPP 5
will set:
•
user port pins OUT0 and OUT2 to high level
•
user port pins OUT1 to low level
Содержание NARDA PMM ER8000
Страница 28: ...2 10 Installation This page has been left blank intentionally ...
Страница 38: ...4 4 Applications Fig 4 2 ER8000 internal LISN Fig 4 3 ER8000 with external L2 16B LISN ...
Страница 40: ...4 6 Applications This page has been left blank intentionally ...
Страница 52: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 82: ...Suggerimenti Commenti Note Suggestions Comments Note ...