Remote control
7-9
SADT b
This setting command sets the detector for Analyzer Mode. The argument (b) should
be a index representing the detetector as follows:
o
1
Peak
o
2
Avg
o
3
Rms
The reply is
SADT =OK
which acknowledges the command has been granted or
SADT =SERR
if the command has been ignored.
Example:
# SADT 1*
sets the detector to Peak for Analyzer Mode
SAFF start,stop
This setting command sets both Start and Stop frequency for Analyzer mode. The
strings (Start and Stop) can be in exponential form and should be expressed in Hz.
The reply is
AFF=OK
which acknowledges the command has been granted or
AFF
=SERR
if the command has been ignored.
Example:
#SAFF 150e3,30e6*
Sets the Start frequency to 150kHz and Stop
frequency to 30MHz for Analyzer mode
SAGO
This setting command puts the ER8000 in Analyzer mode. The reply is always
SAGO=OK,
which acknowledges the command has been granted, followed by an
array, variable in size, representing the reply. Please refer to
“Analyzer Reply
”
section for details of the response.
The ER8000 goes to Analyzer mode.
The ER8000 goes to Analyzer mode.
Example:
#SAGO*
SAHT h
This setting command sets hold time for Analyzer mode. The string (h) should be
expressed in ms. The reply is
AHT=OK
which acknowledges the command has been
granted or
AHT =SERR
if the command has been ignored.
Example:
#S AHT 1500*
Sets the hold time, for Analyzer mode, to 1500 ms (or 1.5s)
SAOP f
This setting command sets the Stop frequency for Analyzer mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is
AOP=OK
which
acknowledges the command has been granted or
AOP =SERR
if the command has
been ignored.
Example:
#SAOP 10e6*
Sets the Stop frequency, for Analyzer mode, to 10MHz
SART f
This setting command sets the Start frequency for Analyzer mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is
ART=OK
which
acknowledges the command has been granted or
ART =SERR
if the command has
been ignored.
Example:
#SART 150e3*
Sets the Start frequency, for Analyzer mode, to 150kHz
SCFA
This setting command
de
activates
t
he conversion factor (if any).
The reply is:
CFA=OK (OFF)
Содержание NARDA PMM ER8000
Страница 28: ...2 10 Installation This page has been left blank intentionally ...
Страница 38: ...4 4 Applications Fig 4 2 ER8000 internal LISN Fig 4 3 ER8000 with external L2 16B LISN ...
Страница 40: ...4 6 Applications This page has been left blank intentionally ...
Страница 52: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 82: ...Suggerimenti Commenti Note Suggestions Comments Note ...