Getting Started
Introduction and Safety Precautions
The Benchmark C Heated Laboratory Centrifuge is intended for use in Class I, Division 2 testing
laboratories and refineries. It operates on 120/240VAC and comes completely assembled and ready
for installation. Also included are a set of tubes, cushions, shields and collars (four each). Unpacking
and lifting the centrifuge should always be done by at least two people. Retain the custom packaging
materials for future shipping or storage.
The unit is equipped with a programmable digital temperature controller with an LCD display. The
RTD meter, or temperature controller, is preset at 140°F (60°C). There is an encoder driven
tachometer with an LCD display of the actual motor speed. Variable speed motor control is included
with a preprogrammed acceleration/deceleration ramp to ensure smooth operation.
The programmable countdown timer is integrated with the motor controller – the countdown begins
only upon reaching the designated speed for the test. Both the timer and the temperature controller
settings are adjustable.
This centrifuge is designed to optimize operator safety and includes an automatic-locking door.
Carefully following safety guidelines (i.e. OSHA) and knowing the limits of this unit will help prevent
accidents.
REMOVE
all foreign objects from the centrifuge bowl.
Keep the sliding door
CLOSED
, especially while samples are spinning.
NEVER
try to slow the trunnion arm down manually. Allow the trunnion arm to come to a
complete stop before removing samples.
Turn all switches
OFF
when not in use.
NEVER
leave the centrifuge unattended while it is turned on or operating.
DISCONNECT
the power supply before removing or replacing electrical or mechanical parts.
DO NOT
leave oily or solvent saturated rags in or around the unit.
DO NOT
allow unauthorized persons to operate the centrifuge.
BE AWARE
of surroundings.
DO NOT
operate the unit when fatigued or under the influence of
medication, alcohol or illegal substances.
If a tube has broken inside of a shield, the cushion
MUST
be replaced.
Use
CAUTION
during heated operation, as the top of the unit is hot to the touch.
The following guidelines can help prevent user error/injury and improve operation: