42
(11) [EIC]
Contents: Conducts
sensitivity registering calibration.
(Relevant to panel operation F-03)
Format: EIC0x.xxx,
±yyyyyy<CR>
Parameter: 0x.xxx, ±yyyyyy
Indicates the indicated value with a decimal point.
If the indicated value has no decimal point, it is
filled with ‘0’ from the head.
Sensor rated output value (mV/V)
Return value format:
ok<CR><LF>
When error occurs, E-00<CR><LF>
Description & others:
Saves the results of this command.
(12)
[EID]
Contents:
Automatically
conducts sensitivity registering calibration with TEDS
information.
Format: EID<CR>
Parameter: None
Return value format:
ok<CR><LF>
When error occurs, E-03<CR><LF>
Description & others:
Among TEDS information, rated capacity, rated output, and recommended
excitation voltage are read to set sensitivity registration and excitation
voltage.
(13) [SMF]
Contents:
Set the hold mode. (Relevant to panel operation F-04.)
Format: SMFa,
b,c,x.xx,yy.yy<CR>
Parameter: a, b, c,x.xx,yy.yy
Specified time (sec)
Delay time (sec)
Comparison mode
0: Normally compare 1: Compare at hold mode
Display mode
0: Normally display
1: Display at hold mode
Hold mode
0: Tracking
1: Arbitrary point hold
2: Peak hold
3: Block specified peak hold
4: Time specified peak hold
Return value format:
ok<CR><LF>
Description & others:
Corresponding to ‘/M’ command