13
Function
Selecting Mode
Function Name
Key operation & WGA-670B
Movement in Function Setting Mode
Functional
Description
F-02
Actual Load
Calibration
●
The indication ‘00628’ or ‘11111’ is a
key code to prevent calibration value
from carelessly changing due to
malfunction.
F-03
Sensitivity
Registering
Calibration
Set transducer rated output (Unit: mV/V)
and press the SET key.
Next, set the required value to be
indicated and press the SET key.
Then, set a decimal point position and
press the SET key.
•Key code enable setting
‘00000’ is indicated.
Here, set the indication to read ‘00628’ or
‘11111’ and press the SET key. Then,
calibration value is updated.
The key code is enabled. Therefore, when
the calibration value is updated the next
time, the above operation is required. To
disable the key code, use the function
‘F-12.’
•Key code disable setting
‘00000’ is indicated. Press the SET key
as it is. Then, the calibration value is
updated.
The key code is disabled. The calibration
value is updated even if no key code is
entered.
●
The indication ‘00628’ or ‘11111’ is a
key code to prevent calibration value
from carelessly changing due to
malfunction.
●
Setting available range is ‘10000’ or
less per 0.1 mV/V.
•A function that
conducts calibra-
tion by registering
the rated output
and rated capacity
of the transducer
without applying
the actual load.
•Factory default is
‘4000’ when 2.000
mV/V is entered.
•For details, see
“4-2-4.”