4. Connection 

Caution

The cable  tip  of  the  product  is assembled  with  the  Lead-free  solder.  Be 

sure  to  use  the  Lead-free  solder  (Sn96.5%,  Ag3%,  Cu0.5%)  when 

soldering the cable.

4.1 Connect the transducer to strain amplifier, using the accessory cable. 

4.2 When using an NDIS connector (PRC03-12A10-7M), perform wiring as illustrated 

below. 

4.3 How to Connect When Using KYOWA’s Bridge Box 

4.4 It is required to heat-run for 5 to 10 minutes before starting the measurement. 

4.5 A  shield wire is not attached  to the LMA-A. If  necessary, make grounding at the 

amplified  indicator.  At  this  time,  if  inductive  noise  is  generated,  connect  an 

oscilloscope to the output terminal of the amplified indicator, and while observing 

the waveform presented, perform suitable grounding. Note that grounding does not 

always produce a good result. 

4.6 Combining with measuring instrument, Measuring Method of Data Logger setting 

CH  mode  (ex.  UCAM-60B  and  UCAM-500)  is  "4-gage  method  (Strain  gage 

transducer) ". 

Select the CH mode from "4-Gage (Constant voltage)". 

5. Conversion 

5.1 Use the rated output described in the Test Data Sheet to convent a reading into a 

load value. 

5.2  When  a  strain  amplifier  is in  use,  output reads  in  m/m  equivalent strain  (×10

-6

equivalent strain). Use the following equation to obtain a load value. 

Load Value(N)

Strain amplifier output ( m/m)

×

Rated Capacity (N)

Rated output (mV/V)

2000( m/m

mV/V)

5.3  When  using  an  amplifier  of  other  type  or  a  recorder,  first  find  the  exact  bridge 

exciting voltage applied. The Test Data Sheet states the rated output caused by a 

bridge excitation voltage of 1V. Use the following equation to obtain a load value. 

Load Value(N)

Bridge output voltage(mV)

×

Rated Capacity (N)

Bridge excitation voltage( V)

Rated output (mV/V)

Sensitivity Decrease due to Cable Extension 

If  a  strain  gage  transducer  is  connected  to  a  signal  conditioner,  digital  indicator  or 

strain amplifier via extension cable, we cannot ignore the sensitivity decrease due to 

the extension cable resistance which lowers the voltage applied to the transducer. 

The  rated  output  with  lowered  sensitivity  can  be  obtained  through  the  following 

equation; 

[

ε

0

 ] = [ R ] × [

ε

i

 ] ÷ [ R

( r × L ) ] 

R

:

 Transducer’s input resistance (

:

 Extension cable’s reciprocating resistance (

) per meter 

L

:

 Extension cable length (m) 

ε

i

:

 Rated output written in Test Data Sheet. 

6. Maintenance and inspection 

6.1 Avoid water, dust and oil on the end of the cable. 

6.2 If a suspicious initial value  or reading appears, measure input resistance, output 

resistance as well as insulation resistance (which should be 100M

 or higher). If 

abnormal resistance is found, the cause may be failure of the sensing element. 

In this case, contact your KYOWA representative for necessary inspection. 

Caution

To measure insulation resistance, apply a voltage lower than 50V to the insulation 

resistance tester.

7. Specifications 

Model

Rated Capacity

Natural frequency

LMA-A-5N 

5N (509.9gf) 

Approx. 15.3kHz 

LMA-A-10N

10N (1.020kgf)

Approx. 17.5kHz

LMA-A-20N 

20N (2.039kgf) 

Approx. 24.8kHz 

LMA-A-50N

50N (5.099kgf)

Approx. 32.6kHz

LMA-A-100N 

100N (10.20kgf) 

Approx. 21.6kHz 

LMA-A-200N

200N (20.39kgf)

Approx. 29.7kHz

LMA-A-500N 

500N (50.99kgf) 

Approx. 43.9kHz 

LMA-A-1KN

1kN (102.0kgf)

Approx. 53.0kHz

NOTE:

 The units and numerical values in parentheses in the rated capacity 

column depend on the conventional unit expression just for reference. 

Safe Overload Rating 150

Rated output 0.6 to 2mV/V(1200 to 4000 m/m)  [5N] 

0.75 to 2mV/V(1500 to 4000 m/m) [10N to 1KN] 

Nonlinearity Within ±1%RO 

Hysteresis Within ±1%RO 

Recommended excitation voltage 1 to 5V AC or DC 

Safe excitation voltage 7V AC or DC 

Input resistance 350 ±2.5% 

Output resistance 350 ±2.5% 

Repeatability 1%RO MAX 

Compensated temperature range 0 to 50

(noncondensing) 

Safe temperature range

  −

10 to 60

(noncondensing) 

Temperature effect on Zero balance Within ±0.3%RO/

    [5N] 

Within ±0.2%RO/

    [10N to 50N] 

Within ±0.05%RO/

  [100N to 1KN] 

Temperature effect on output Within ±0.2%/

    [5N to 50N] 

Within ±0.05%/

  [100N to 1KN] 

Cable 4-conductor (0.035mm

2

) vinyl shielded cable,   

1.7mm diameter by 2m long, bared at the tip 

(Shield wire is not connected to mainframe) 

Weight    Approx. 1.5 g (cable not included) [5N to 50N] 

Approx. 11 g (cable not included) [100N to 1KN] 

Accessories Instruction manual 1 

                        Test Data Sheet      1 

Warranty

      1 

                        RoHS Directive    2011/65/EU 

Harmonised standards    EN50581:2012 

Transducer