6-4 Diode Measurement
(1) Set the Function Selector Switch to "
".
(2) Press the
SEL
Button. "
" mark is shown on the display.
(3) Connect the test leads to the circuit under test.
The measured value appears on the display.
[ Forward-bias Diode Test ]
Connect the red test lead to the anode and the black test
lead to the cathode.
[ Reverse-bias Diode Test ]
Connect the red test lead to the cathode and the black test
lead to the anode.
Note:
◇
Hold the clamp sensor in the holster, while making the
diode measurement
6-5 Continuity Measurement
(1) Set the Function Selector Switch to "
".
(2) Press the
SEL
Button. " " mark is shown on the display.
(3) Connect the test leads to the circuit under test.
The measured value appears on the display.
When the measurement value is lower than 35±25
Ω
, the
instrument beeps.
Note:
◇
Hold the clamp sensor in the holster, while making the
continuity measurement
6-6 Capacitor Measurement
(1) Set the Function Selector Switch to "
".
(2) Press the
SEL
Button. "F" mark is shown on the display.
(3) Connect the test leads to the circuit under test.
The measured value appears on the display.
Note:
◇
Hold the clamp sensor in the holster, while making the
capacitor measurement.
6-7 Frequency Measurement
#
DANGER
●
In order to avoid possible shock hazard, never make
measurement on circuits with a maximum voltage difference
of 600V or greater between conductors (300V or greater
between a conductor and ground).
●
Do not make measurement with the test leads connected
to the circuit under test. Never make measurement with the
battery compartment cover removed.
●
Do not make current measurement with the test leads
connected to the circuit under test.
●
Keep your fingers and hands behind the protective fingerguard
during measurement.
(1) Set the Function Selector Switch to "Hz."
(2) Measuring frequency of current:
Adjust one of the conductors to the center of the clamp
sensor's arrow.
Measured value is shown on the display.
Measuring frequency of voltage:
Connect the test leads to the circuit under test. Measured
frequency is shown on the display.
Note:
◇
Do not use both the clamp sensor and the test leads at
the same time, while making the frequency measurement.
◇
Hold the test leads in the holster, while making the frequency
measurement by using the clamp sensor.
◇
Hold the clamp sensor in the holster, while making the
frequency measurement by using the test leads.
Black
Red
Forward-bias Diode Test
Test Lead
Anode
Cathode
Red
Black
Reverse-bias Diode Test
Test Lead
Anode
Cathode
Prohibition