6-4 Diode Measurement

(1) Set the Function Selector Switch to " 

 ".

(2) Press the 

SEL

 Button. "

" mark is shown on the display.

(3) Connect the test leads to the circuit under test.

     The measured value appears on the display.

     [ Forward-bias Diode Test ]

         Connect the red test lead to the anode and the black test 

lead to the cathode.

     [ Reverse-bias Diode Test ]

         Connect the red test lead to the cathode and the black test 

lead to the anode.

Note:

 Hold the clamp sensor in the holster, while making the 

diode measurement

6-5 Continuity Measurement

(1) Set the Function Selector Switch to " 

 ".

(2) Press the 

SEL

 Button. "   " mark is shown on the display.

(3) Connect the test leads to the circuit under test.

     The measured value appears on the display.

      When the measurement value is lower than 35±25

Ω

, the 

instrument beeps.

Note:

 Hold the clamp sensor in the holster, while making the 

continuity measurement

6-6 Capacitor Measurement

(1) Set the Function Selector Switch to " 

 ".

(2) Press the 

SEL

 Button. "F" mark is shown on the display.

(3) Connect the test leads to the circuit under test.

     The measured value appears on the display.

Note:

 Hold the clamp sensor in the holster, while making the 

capacitor measurement.

6-7 Frequency Measurement

#

 

DANGER

 In order to avoid possible shock hazard, never make 

measurement on circuits with a maximum voltage difference 

of 600V or greater between conductors (300V or greater 

between a conductor and ground).

 Do not make measurement with the test leads connected 

to the circuit under test. Never make measurement with the 

battery compartment cover removed.

 Do not make current measurement with the test leads 

connected to the circuit under test.

 Keep your fingers and hands behind the protective fingerguard 

during measurement.

(1) Set the Function Selector Switch to "Hz."

(2) Measuring frequency of current:

      Adjust one of the conductors to the center of the clamp 

sensor's arrow.

     Measured value is shown on the display.

     Measuring frequency of voltage:

      Connect the test leads to the circuit under test. Measured 

frequency is shown on the display.

Note:

 Do not use both the clamp sensor and the test leads at 

the same time, while making the frequency measurement. 

 Hold the test leads in the holster, while making the frequency 

measurement by using the clamp sensor. 

 Hold the clamp sensor in the holster, while making the 

frequency measurement by using the test leads.

Black

Red

Forward-bias Diode Test

Test Lead

Anode

Cathode

Red

Black

Reverse-bias Diode Test

Test Lead

Anode

Cathode

Prohibition

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание KEW MATE 2012RA

Страница 1: ...DIGITAL MULTIMETER WITH AC DC CLAMP SENSOR KEW MATE 2012RA www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...his instrument has been designed and tested according to IEC Publication 61010 Safety Requirements for Electronic Measuring Apparatus This instruction manual contains warnings and safety rules which m...

Страница 3: ...omes standard with the instrument Clamp sensor for ease of use in crowded cable areas and other tight places Permits current measurement with an open current clamp sensor that does not require opening...

Страница 4: ...Input Voltage 600V Range Display range Allowable input Accuracy 600mV 0 0 603 9mV 0 0m 600 0V 1 0 rdg 3dgt 6V 0 560 6 039V 60V 5 60 60 39V 600V 56 0 603 9V Input Impedance approx 10M Resistance Auto...

Страница 5: ...g System modulation Display Liquid crystal display Maximum Reading 6039 Except Hz 9999 CAP 4039 Diode 1999 Bar graph with maximum points of 30 Display approx 3 times per second Indication renewal Loca...

Страница 6: ...the batteries according to section 8 Battery Replacement NOTE When the instrument is left powered on the auto power save function automatically shut the power off The display blanks even if the Functi...

Страница 7: ...LCD 2 Adjust one of the conductors to the center of the clamp sensor s arrow When the position of the conductor is not at the center of the arrow the error occurs Measured value is shown on the displa...

Страница 8: ...urement DANGER In order to avoid possible shock hazard never make measurement on circuits with a maximum voltage difference of 600V or greater between conductors 300V or greater between a conductor an...

Страница 9: ...Turn the function switch to the OFF and then to any position 7 2 Data Hold Function This is a function to freeze a measured value on the display Press the Button once to hold the current reading In t...

Отзывы: