![KYORITSU KEW 1051 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/kyoritsu/kew-1051/kew-1051_instrucion-manual_2017125034.webp)
33
5. User Calibration Function
It is recommended that the instrument be calibrated periodically.
The instrument can be calibrated.
CAUTION
To avoid electrical shock
• Only authorized engineers are allowed to calibrate the instrument using dedicated facilities.
• Connect the calibrator to the instrument with the calibrator’s test leads.
• Before carrying out calibration, read the instruction manual of the calibrator.
• Temporarily remove the test leads from the instrument before switching measurement mode
(function).
<Conditions of calibration>
Calibrator: With accuracy higher than of this instrument
Ambient Environment:
Temperature: 23±3ºC
Humidity: 55%RH or less
Leave the instrument for 30 minutes under above conditions
before carrying out calibration.
After reference valve of Calibrator stabilizes, Press the key to confirm for calibration valve.
Содержание KEW 1051
Страница 1: ...INSTRUCTION MANUAL DIGITAL MULTIMETER KEW 1051 1052...
Страница 39: ......