![Kyoritsu Electrical Instruments Works 6200 Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/kyoritsu-electrical-instruments-works/6200/6200_instruction-manual_2017191012.webp)
―
10
―
Measurement of current consumption (ILINE) (Only for M-6202)
Threshold and display (Threshold display is dependent on the accuracy on each
measuring range.)
Reference conditions
Unless otherwise specified, this specification is dependent on following
condition.
(1) Ambient temperature
:
23
±
5
℃
(2) Relative humidity
:
45 ~ 75%
(3) Position
:
Horizontal
(4) AC power source
:
230V, 50/60Hz
(5) Altitude
:
2000m or less
Operating temperature and humidity range
0
℃
~ +40
℃
Relative humidity: 85% or less (no condensation)
Storage temperature and humidity range
-20
℃
~ +60
℃
Relative humidity: 85% or less (no condensation)
Rate voltage and frequency
Rated voltage
:
230V
±
10%
Rated frequency
:
50 ~ 60Hz
±
1%
Operating method
:
Dual slope integration
Display
:
3 1/2 digit in LCD, character height 17mm,
Max reading 1999 with units, symbols, etc.
Maximum rated power
Approx. 9VA
Maximum throughput of M-6202
Max. 4000VA
Rating
Measuring range
Resolution
Accuracy
Throughput
16A AC
0 ~ 15.99A AC
10mA
±
(2% rdg+3dgt)
Max. 4000VA
Range
Protective conductor resistance
Insulation resistance
Equivalent leakage current
Contact current
Threshold
LED for threshold
display
Warning buzzer
(continuous sound)
RPE >0.3
Ω
RPE >1
Ω
RINS < 0.5M
Ω
RINS < 2.0M
Ω
IEL >7.0mA
IEL >15mA
IL >0.25mA
IL >0.5mA
●
●
●
●
●
●
●
●
---
●
●
---
---
●
---
●