3.6 Explanation for indications
3.6-1 M-6200 Display (common with M-6202)
Note. 1) Over range display: "OL" symbol is displayed on the LCD over range
display.
2) LINE = PHASE
3.6-2 LED for threshold value indication and buzzer
When exceeding following limit values which are defined in DIN
VDE0702:1995, LED lights up and buzzer sounds.
3.7 Applicable standards
Instrument operation: DIN VDE 0701:2000 / DIN VDE 0702:1995
Safety: IEC/EN61010-1 CATIII 300V-instrument
IEC/EN61010-2-031 CATIII 300V(600V)-test lead
EMC Interference Emission / Immunity : EN 61326-1
As for basic standards, EN 61000-4-3, EN 61000-4-4, EN 61000-4-6, and
EN 61000-4-11, criterion B.
(Criterion B: Temporary degradation of performance or loss, but self-recovering)
―
8
―
Measuring range
Condition
LED for threshold
value indication
Warning buzzer
(continuous sound)
RPE >0.3
Ω
RPE >1
Ω
RINS < 0.5M
Ω
RINS < 2.0M
Ω
IEA >7.0mA
IEA >15mA
IL >0.25mA
IL >0.5mA
●
●
●
●
●
●
●
●
---
●
●
---
---
●
---
●
Protective conductor
resistance
Insulation resistance
Equivalent leakage
current
Leakage current /
contact current
(substantiation of
absence of voltage)
Fig.6
The Polarity of protective
earth conductor resistance
Power supply voltage
(Line-Neutral)
Current consumption
NULL indication
Units