Rear View
1.
Mains power.
Use supplied Neutrik Powercon power cable to connect the laser system to mains
power supply.
2.
Mains power switch.
3.
Safety key switch.
The Safety key must be inserted and turned to ON position in order to enable
the system to operate.
4.
E-STOP Remote INPUT/OUTPUT.
In order to use the laser system, the Interlock must be closed
circuit. This is done by connecting the Emergency STOP Remote to the XLR Remote INPUT
socket on the projector using the supplied cable and plugging in the User Interlock Connector into
the Remote’s XLR “User E-STOP Connector”.
The User Interlock Connector may be replaced by the user’s own interlock system using a switch
or dry relay closure to connect pins 1 and 2. With pins 1 and 2 shorted, Laser Emission is
possible, provided all other interlocks are closed circuit. With pins 1 and 2 open, NO Laser
Emission is possible.
The projector’s
REMOTE Output
is used to daisy-chain the Remote signal when you want to use
a single Emergency STOP Remote to control multiple systems.