117
117
117
117
frequency measurements return “0”.
CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure a period measurement with the specified range and
resolution. This command does not initiate the measurement. For period
measurements, the meter uses only one “range” for all inputs between 0.33
seconds and 3.3 µsec. With no input signal applied, period measurements return “0”.
CONFigure:CAPacitance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for capacitance measurements with the
specified range and resolution. This command does not initiate the
measurement.
CONFigure:CONTinuity
Preset and configure for a continuity measurement. This command does not
initiate the measurement. The range and resolution are fixed at 1kΩ and 5½
digits respectively.
CONFigure:DIODe
Preset and configure for a diode measurement. This command does not initiate
the measurement. The range and resolution are fixed at 1Vdc with 1mA current
source and 5½ digits respectively.
CONFigure:TEMPerature
Preset and configure for a RTD measurement. This command does not initiate
the measurement. The resolution is fixed at 6½ digits.
CONFigure:TCOuple
Preset and configure for a thermocouple measurement. This command does not
initiate the measurement. The resolution is fixed at 6½ digits.
CONFigure?
Query a recent condition of function, and return a quoted string.
C.5 The Measurement Configuration Command
[SENSe:]FUNCtion “<
function
>”
Содержание M3500
Страница 6: ......
Страница 10: ......
Страница 12: ...w a r ...
Страница 13: ......
Страница 14: ......
Страница 16: ......
Страница 29: ......
Страница 37: ......
Страница 41: ......
Страница 42: ...In Short Condition ...
Страница 43: ......
Страница 44: ......
Страница 109: ...104 104 104 104 651 Panel Program Checksum Error The checksum from the panel is incorrect ...
Страница 110: ...K U S A M ME C O ...
Страница 148: ...K U S A M M E C O K U S A M M E C O ...
Страница 155: ...K U S A M M E C O ...
Страница 163: ...158 158 158 158 m_nCount 0 m_defaultRM_usbtmc 0 FreeLibrary hUSBTMCLIB hUSBTMCLIB NULL return ...
Страница 164: ......