15

7)  After  setting  the  parameters,  remove  the 
      standard component and insert the component 
        to  be  measured.  If  the  impedance  measured 
     does not exceed tolerance range, the primary 
      display  will  show    "PASS",  otherwise  show 
      "FAIL". The current measurement result will be 
      shown on the secondary display. 

3)   When sorting mode is active, press 

SETUP 

key

        to  modify  the  range,  reference  value  and  the 
      tolerance settings sequentially.

4)  "Range"  symbol  is  flashing  when  setting  the 
      range. Press 

D/Q/ESR

 (     ) key to shift the decimal

        point,  unit  to  left  and  press 

SER/PAL

  (        )  key  to 

        right.  Press 

ENTER 

key  to  confirm  and  enter  into 

        the  reference  value  setting  mode  automatically. At
      this time, "Range" symbol disappear.

5)  When setting the reference value, press 

D/Q/ESR

     (   ) key and 

SER/PAL

 (   ) key to shift the bit to left 

      and right respectively. Press 

PCLINK 

(    ) key and 

       

REL  % 

(      )  key  to  make  the  digit  +1  or  -1.  The 

      flashing bit is the current setting bit. The reference 
     value setting is available from 20 to 1999 counts. 
      Press 

ENTER 

key  to  confirm  and  enter  into  the 

      tolerance setting mode automatically.

6)  When  setting  the  tolerance,  press  

D/Q/ESR

 (   )

     key & 

SER/PAL

    key to select tolerance range : 

     ± 0.25%     ± 0.5%     ±1%     ±2%     ±5%     ±10%       
           ±20%      ± 80%-20%. The default tolerance is 
     ±1%. Press 

ENTER

 key to confirm.

16

8)  Press 

SORTING 

key again to exit the sorting mode.

Note:

 In AUTO LCR mode, the SORTING key is 

not available.

9.  PC-LINK mode

Press 

PCLINK

  key  "USB"  symbol  appear  on 

LCD.  Connect  the  instrument  to  PC  through 
USB  interface,  and  the  measured  data  can  be 
recorded,  analyzed,  processed  and  printed  by 
PC. Press 

PCLINK

 key again to cancel the data 

transmission. Then USB symbol disappears.

Due  to  the  power  consumption  in  data 
transmission, please exit USB mode when there 
is no need to transmit data.

Operating step

1.Inductance measurement

1)  Turn on the power.
2)  There are two ways for inductance input:
a.  Insert  the   inductance  to be measured into the 
     input terminal directly (shown in figure-8);
b.  Connect the alligator clips to the ends of the 
     inductance  to  be measured (shown in figure-9).
3)  The  default  test  mode  is Auto LCR mode, the 
     inductance   value   will  be  shown  on  primary 
     display  and  the  quality  factor  Q will shown on 
     secondary  display. In Auto LCR mode, the 

D/Q/

     ESR 

 key,  

SEL /  PAL 

 key,  

SORTING

 key and 

     

REL%

 key are not available

Содержание KM 520B

Страница 1: ...meter Wood Paper Grain Moisture Meter Distance Meter Network Cable Tester Digital Hand Held Temperature Indicators Digital Lux Meter Thermal Imaging Camera Power Factor Regulator Maximum Demand Contro...

Страница 2: ...o turn off the instrument when replacing the battery or DC power adapter Although the instrument has a protection for shocks inputing dc voltage or current will still damage the instrument The capacit...

Страница 3: ...20nF 20mF 2000pF 2mF 200pF 200m F 200pF 20m F 200W 200MW 20W 200MW 20W 20MW 20W 2MW C R DCR D Q q Test Level Range mode Equivalent circuit 0 6 Vrms Auto and Hold Parallel and Series 4 Specifications C...

Страница 4: ...ity factor ESR equivalent series resistance Rp equivalent series parallel resistance q phase angle 6 Explanation on Front Panel The front panel is shown as figure 1 1 Mini USB interface Connect with P...

Страница 5: ...y When sorting mode is active press SETUP key to modify the reference value range and the tolerance setting sequentially 11 SER PAL key Used to select series and parallel mode 12 FREQ key Press FREQ k...

Страница 6: ...ay is quality factor Q C D Capacitance measurement the parameter on sub display is dissipation factor D R Resistance measurement DCR DC resistance measurement mode L C R measurement readings can be po...

Страница 7: ...bers the current readings on primary display called initial value when pressing the REL key and 12 b When using insert the black and red testing lines with alligator clip into the terminal and termina...

Страница 8: ...gs to the component type press FUNC key to select L C or R measurement mode 2 Insert the standard component into the input therminal Press SORTING key to enter into the sorting mode and the Sorting sy...

Страница 9: ...nce setting mode automatically 6 When setting the tolerance press D Q ESR key SER PAL key to select tolerance range 0 25 0 5 1 2 5 10 20 80 20 The default tolerance is 1 Press ENTER key to confirm 16...

Страница 10: ...x Lx Cx LCUR LPOT HPOT HCUR GUARD GUARD INPUT Rx Lx Cx LCUR LPOT HPOT HCUR Figure 8 Figure 9 18 4 The default test mode is Auto LCR mode the capacitance value will be shown on primary display and the...

Страница 11: ...ilable 3 The default test mode is Auto LCR mode the resistance value will be shown on primary display and the phase angle q will be shown on secondary display In Auto LCR mode the D Q ESR key SEL PAL...

Страница 12: ...t and then firmly insert DC plug into DC input end in the right of the meter When unplugged firstly pull out the DC plug perpendicular to DC input end and then unplug the AC plug from the electrical o...

Страница 13: ...s warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of KUSAM MECO KUSAM MECO s warranty obligation is limited at...

Отзывы: