360208-F    454FTB, 504FTB, 534FTB 

 

 

 

Insertion Meter Installation Guidelines 

1.

 

Mount Probe so its velocity sensing element is centered in the duct/pipe.  This location has the most stable flow reading.  Note the flow arrow direction 
points in the direction of the flow. 

2.

 

The upstream and downstream distance to flow profile disruptions from that of a straight pipe are “X” diameters upstream and 5 downstream.    We have 
chosen 2% maximum error from the baseline straight pipe calibration for the distance criteria.  Longer straight runs will reduce this error level.  Here are 
four example cases in order of their disruption to the flow: 

a.

 

Valves change the flow profile as they open and close so the sensor should not be located too close.  The gate valve is the worst and the metering 
valves are the best at reducing their flow profile changes between open and closed.   

b.

 

Branching joints also change the profile as the percentage of flow changes between the branches. 

c.

 

Elbows or direction changes will disrupt the long-run pipe profile.  As this disruption settles down, the profile can wobble or move depending on 
the flow rate.  Several elbows in different planes will impart a swirl that also imparts an error in the readings (flow is not straight on to the sensor) 
and the upstream distance should be increased.  The distance from elbows may be reduced if field calibrations are used.  Double elbows in different 
planes introduce swirl and need 50% more distance than a single elbow does. 

d.

 

A line size change will also disrupt the profile and can introduce instability so this should be avoided unless you have more experience on how to 
use this to your advantage.   The distance from a line size change may be reduced if field calibrations are used. 

3.

 

The duct or pipe inside dimensions are used to determine the flow area of the meter.  An area wizard in the meter setup menu walks you through entering the 
data and the meter will calculate the flow area. 

4.

 

The probe will block off some of the flow area and accelerates the velocity proportional to its area projected on to the duct/pipe cross section.  This is known 
as the Sensor Blockage Correction Factor (SBCF).  You enter the insertion depth (L), which is measured from the end of the window to the duct inside wall, 
in the meter setup menu and the meter will calculate the SBCF. 

5.

 

Field Calibration Data is the key to achieving accuracy for the insertion probe.  Without this step you can have good repeatability but the absolute flow 
number requires a reference flow measurement be taken at the same time you record the indicated display of the insertion meter.  The correction factor 
which is velocity dependent,  CF(v), is defined as the ratio of True reading to Indicated reading.  The basic calibration of the unit is a point velocity sensor.  
To convert this to volumetric flow rate or mass flow rate requires the proper area and the average velocity, this is where the field calibration fits in.  As the 
field calibration method is quite long and technical, you are encouraged to do one of two things to achieve this: 

a.

 

Contract the field calibration with Kurz Instruments, Inc. or another reputable field calibration company. 

b.

 

Do it yourself if you have the experience or a test team at your disposal. 

 

Common methods of doing reference method calibrations entail: duct traverses, tracer gas dilution method or the process Stoichiometry which requires support 
from your process engineering group.  Sometimes the correction factors can be entered based on an ideal duct profile model or a finite element analysis (FEA) or 
computation fluid dynamics (CFD) program. 
 
For more information on all the above, please refer to the full product manual provided on CD or at our website 

www.kurz-instruments.com

.  Also see “Flow 

Measurements 2

nd

 ed.” Edited by D. W. Spitzer, ISA Press 2001. Chapter on Insertion Flow Measurements. 

Содержание 454FTB

Страница 1: ...struments Inc 831 646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 1 Start Up Guide Models 454FTB 504FTB and 534FTB August 2010 For MFT B Series Firmwa...

Страница 2: ...et You will need to remove the protective shipping covers from the sensor or flow body before installing the unit A thermal meter must have direct contact with the process fluid to work The meter will...

Страница 3: ...depth is adequate to get into the center line where the flow profiles are the most stable The distance from profile disruptions needs to be adhered to for the best repeatability accuracy see the guid...

Страница 4: ...Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 4 X Vavg Vs CF v CF v Vavg Vs 0 5 CF v 1 0 Kurz 4xx 454 etc D A ft2 or m2 A D2 4 d L L ft m d in SBCF A A 12dL 5 D CF v v x x x x 0 7 1 0 Vs 2 X 40D X...

Страница 5: ...s are used to determine the flow area of the meter An area wizard in the meter setup menu walks you through entering the data and the meter will calculate the flow area 4 The probe will block off some...

Страница 6: ...646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 6 X Kurz 5xx 504 etc 5 D Kurz 534 0 D 0 D D X 40D X 20D X 15D 2 Less than two line size changes Less t...

Страница 7: ...tially calibrated using long straight runs with short close coupled pipe flanges that minimize the line size discontinuity In many applications field calibration data can be used to reduce the upstrea...

Страница 8: ...Kurz Instruments Inc 831 646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 8...

Страница 9: ...Kurz Instruments Inc 831 646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 9...

Страница 10: ...F 454FTB 504FTB 534FTB 10 Customer I O Panel AC Power 2 2 or 4 4 20 mA AO FLOW Temp 24 VDC Digital I O 454 Vdata Area CF v Customer I O Panel AC Power 2 2 4 20 mA velocity 24 VDC 454 Vdata 4 20 mA AO...

Страница 11: ...Series assume the PLC is connected to a loop powered device Potentially Hazardous Area Use and EMI Compliance Please consult the full manual on the provided CD for safety and EMI related issues Sensor...

Страница 12: ...age for the meter s AO interface 2 You can self power the 4 20 mA circuit from the flow meter where Kurz provides the 24 VDC excitation voltage The output is not isolated in this case Here your 4 20 m...

Страница 13: ...long enough if not twisted up too much to rotate into any of the four positions Navigating Menus The MFT B flow meter with 2 0 and later firmware has a new menu system that is different than the 1 x...

Страница 14: ...e changed by the user A Quick Look up Card for the B Series 2 0 is provided with each meter The Quick Lookup Card is a handy navigational aid for the B Series 2 0 menu system Additional cards are avai...

Страница 15: ...TAG NAME KURZ MFT B The TAG NAME is entered using the numeric keys on the keypad 0 9 similar to entering a text message on a cellular phone Alphabetic characters and special characters e g etc are ma...

Страница 16: ...its for Velocity and Temperature Velocity SFPM SMPS NMPS Temperature DEGF DEGC The duct pipe area blocked off by the insertion of the sensor probe accelerates the flow reading This is adjusted for wit...

Страница 17: ...sions are inside dimensions of the duct DUCT WIDTH 36 459 IN DUCT HEIGHT 40 989 IN Use the numeric keys to enter the value in inches or millimeters depending on the flow units specified previously Pre...

Страница 18: ...ction displayed In this example FLOW RATE is selected for the Channel 1 AO so the following example screens will prompt for the low and high range of the flow rate being measured The next prompt is fo...

Страница 19: ...IC format is selected for the Run Mode Display the meter will prompt for the process variables to display STATIC VARS FLOW TEMP v Use the or v keys to select the desired display option and press E to...

Страница 20: ...LOW VEL FLOW TOT VEL FLOW TEMP FLOW TEMP VEL For the Run Mode scrolled format each screen will display the value of only one process variable the description will be on line 1 and the value with units...

Страница 21: ...d SAVING CONFIG DATA TO EEPROM Data saved to EEPROM is persistent between power cycles of the flow meter Meter Programming with the USB Port The Quick Setup can also be performed using a PC laptop run...

Страница 22: ...and the cursor position is advanced one position to the right for the next character The period and hyphen characters can also be entered in a similar fashion When text input is done press ENTER at th...

Страница 23: ...flow while monitoring the indicated flow from the meter This is best left to experienced users or professionals who know how to do it See the manual for more information on Field Calibration The 2 xx...

Страница 24: ...point protocol that can be used on the RS 485 port Both the Modbus ASCII or RTU protocols are supported The device address may be any number between 1 and 247 Modbus TCP IP requires a serial to Ether...

Страница 25: ...rs and commands field calibration correction factors up to 8 velocity points zero mid span drift check diagnostics alarms sensor calibration purge control damping STP conditions totalizer reset and la...

Отзывы: