360208-F 454FTB, 504FTB, 534FTB
18
For Insertion Meters, the next screen will prompt for the
PROBE DEPTH
PROBE DEPTH
>18.230 IN
Use the numeric keys to enter the value in inches or millimeters depending on the system
units and press
E
to accept the value.
The meter next prompts for the output function for the 4-20mA output channel #1.
ANALOG OUT 1
>FLOW RATE ^v
Use the
^
or
v
keys to display the list of Analog Output selections. The options are
FLOW RATE/VELOCITY/TEMPERATURE/PID. Press
E
to accept the AO1 function
displayed. In this example, FLOW RATE is selected for the Channel 1 AO so the
following example screens will prompt for the low and high range of the flow rate being
measured.
The next prompt is for the analog output minimum range value (the analog output value
at 4mA)
AO1 at 4mA
>0.000000 SCFM
The units depend on the
FLOW UNITS
configured previously. Use the numeric keys to
enter the equivalent flow value at 4mA and press
E
to accept the value.
The next prompt is for the analog output maximum range value (the analog output value
at 20mA)
AO1 at 20mA
>15000.000 SCFM
Use the numeric keys to enter the equivalent flow value at 20mA and press
E
to accept
the value.
If the meter is configured with two analog output channels, the meter will repeat the
analog output setup prompts for AO Channel #2.