Kurz
B
‐
Series
Operations
Guide
4–35
Program Mode
Advanced Setup
Option 15 — Variable Flow Correction
This
is
used
to
correct
the
flow
calibration
data
for
changing
flow
profiles
over
the
entire
velocity
range.
You
can
have
up
to
eight
correction
points
on
a
data
curve.
To
access
the
Variable
Flow
Correction
menu
in
Program
mode:
1>
Press
P
.
2>
Enter
the
Advanced
access
code,
and
then
press
E
.
3>
Press
2
to
invoke
the
Quick
Jump
option.
4>
Press
15
for
the
Variable
Flow
Correction
menu,
and
then
press
E
.
ENTER # OF FLOW
DATA SET >2
The
prompt
for
entering
data
points
appears.
5>
Use
the
number
keys
to
change
the
value
and
then
press
E
.
ENTER FLOW DATA
SET # >1
The
Set
1
prompt
appears.
6>
Press
E
.
REF VALUE R1
>1.200000000 SMPS
The
current
reference
value
for
R1
appears.
It
is
based
on
the
process
reference.
7>
Use
the
number
and
decimal
keys
to
enter
the
reference
value,
and
then
press
E
.
METER DATA D1
>0.66700000 SMPS
The
current
meters
value
for
D1
appears.
This
is
the
value
measured
by
the
flow
meter.
8>
Use
the
number
and
decimal
keys
to
enter
the
flow
meter
value,
and
then
press
E
.
REF VALUE R2
>12.20000000 SMPS
The
Set
2
prompt
appears,
and
the
prompts
repeat
for
each
additional
set.
When
complete,
the
changes
are
saved
and
operations
continue.
Содержание 410FTB
Страница 1: ...Operations Guide for B Series Flow Meters 368042E...
Страница 4: ...Kurz B Series Operations Guide iv...
Страница 9: ...Kurz B Series Operations Guide ix Preface...
Страница 33: ...Kurz B Series Operations Guide 1 20 B Series Flow Meter Modes Menus...
Страница 65: ...Kurz B Series Operations Guide 2 32 B Series Flow Meter Display Mode Options...
Страница 141: ...Kurz B Series Operations Guide 4 68 Program Mode Advanced Setup...
Страница 165: ...Kurz B Series Operations Guide Index 6...