Rotor Name
Nominal Capacity
Tube size (mm)
diam. x length
Maximum Speed
rpm
Maximum RCF
x
g
Tube Rack Adaptor
Code No.
AF-2018
18 x 1.5/2mL
9.5〜11×36〜42
15,000
20,630
—
18 x 0.5mL
7〜7.9×25〜34
15,000
18,110
055-7590
AF-2724A
24 x 1.5/2mL
9.5〜11×36〜42
16,000
23,470
—
12 x 0.5mL
7〜7.9×25〜34
16,000
19,180
—
12 x 0.2mL PCR
5.9〜6.3×20〜24
16,000
16,890
—
AF-2730
30 x 1.5/2mL
9.5〜11×36〜42
14,000
20,820
—
14 x 0.5mL
7〜7.9×25〜34
14,000
17,530
—
14 x 0.2mL PCR
5.9〜6.3×20〜24
14,000
16,430
—
AF-2236
36 x 1.5/2mL
9.5〜11×36〜42
15,000(14,000)
21,880(19,060)
—
12 x 0.2mL PCR
5.9〜6.3×20〜24
15,000(14,000)
15,340(13,370)
—
36 x 0.5mL
7〜7.9×25〜34
15,000(14,000)
19,620(17,090)
055-7590
AF-2048A
48 x 1.5/2mL
9.5〜11×36〜42
12,000(11,500)
15,780(14,490)
—
48 x 0.5mL
7〜7.9×25〜34
12,000(11,500)
14,330(13,160)
055-7590
*2
*2
*2
*2
AF-2018
18 x 1.5/2mL
・
Quick acceleration within 10
seconds
AF-2048A
48 x 1.5/2mL
・
Useful for centrifuging a
large number of samples.
Micro Rotor
Table of Rotors, Buckets and Tube Racks for Model 6200
*
Autoclavable
*
Autoclavable
*
Autoclavable
*
Autoclavable
*
Autoclavable
AF-2730 and AF-5004CS rotors are certified as “Biosealed rotors”
by CAMR (U.K.).
Even if tubes get broken, samples will not come out of the rotor.
These rotors can be used for inspections of BSE, SARS, etc.
AF-2724A
24 x 1.5/2mL, 12 x 0.5/0.2mL PCR
・
4 types of tubes (1.5/2mL,
0.5mL, 0.2mL PCR) can be
used without adaptors.
・
RCF with 20,000 xg is
effective for BSE tests.
AF-2236
36 x 1.5/2mL, 12 x 0.2mL PCR
・
4 types of tubes (1.5/2mL, 0.5mL, 0.2mL PCR)
can be used without adaptors.
・
Caps of 1.5/2mL tubes can be set in the
same direction, which is useful for
experiments.
AF-2730
Biosealed Rotor
30 x 1.5/2mL
・
4 types of tubes (1.5/2mL,
0.5mL, 0.2mL PCR) can be
used without adaptors.
For RCF indication, please refer to the instruction manual.
◉
The temperature of sample is maintained at 4℃ at maximum speed of rotation.
When AF-2236 or AF-2048 rotor is used, The temperature of sample is
maintained at 4℃ at speed of rotation indicated in the brackets ( ), and at 8℃
at maximum speed.
*1
*1
*1
*1
*1
AF-512
12 x 5mL
AF-5004CH
4 x 50mL Conical tube
・
15,000rpm/22,140xg
・
Protective structure for
50mL conical tube caps
prevents the tubes
from getting broken.
(patented)
AF-5004CS
Biosealed Rotor
4 x 50mL Conical tube
・
Various tubes can be
used with adaptors.
AF-5004CA
4 x 50mL Conical tube
AF-5004RA
4 x 50mL
・
10mL biohazard-proof
with sealing adaptors
(code no. 055-4280) can
be used.
AF-3806A
6 x 38mL
Rotor Name
Nominal Capacity
Tube size (mm)
diam. x length
Maximum Speed
rpm
Maximum RCF
×
g
Tube Rack Adaptor
Code No.
AF-5004CH
4 x 50mL conical tube
30 x 117
15,000
22,140
—
4 x 15mL conical tube
17 x 121
15,000
21,130
055-1250
AF-5004CS
4 x 50mL conical tube
30 x 117
12,000
14,970
—
4 x 15mL conical tube
17 x 121
12,000
14,490
055-1280
AF-5004CA
4 x 50mL conical tube
30 x 117
12,000
14,170
—
4 x 15mL conical tube
17 x 121
12,000
13,520
055-1280
AF-5008C
8 x 50mL conical tube
30 x 117
10,000(9,500)
13,420(12,110)
—
8 x 15mL conical tube
17 x 121
10,000(9,500)
13,190(11,910)
055-1280
AF-512
12 x 5mL
15 x 59
15,000
20,880
—
AF-1408A
8 x 14mL
17 x 75〜100
12,000
14,330
—
AF-5004RA
4 x 50mL
28.5〜29 x 102〜115
15,000(14,000)
21,880(19,060)
—
4 x 12mL
15.8〜16.4 x 103〜110
15,000(14,000)
20,630(17,970)
055-7410
4 x 10mL
16〜16.3 x 79〜92
15,000(14,000)
20,630(17,970)
055-4280
AF-3806A
6 x 38mL
25〜25.6 x 87〜103
15,000(14,500)
21,880(19,060)
—
*1 0.2mL PCR tubes can be used with adaptors (code no.055-0500).
When 055-0500 adaptor is used, it could not be used at maximum speed.
*2 This is a speed of outside tubes. Please contact us for more details.
*3 Iwaki, Greiner, Corning, Sarstedt, Sumiron, Vioramo, BD falcon, Fisher, Labcon and Nunc's tubes (cap diam. 36mm or less) can be used.
*4 Nalgene tubes (catalogue no.3110-0500/PPCO, 3110-9500/PPCO, 3117-9500/PC) require adaptors (code no.055-1110).
*5 Biobic 5mL sampling tubes and Watson 5mL mega tubes are applicable.
*6 Falcon tubes (catalogue no.352059) can be used.
*7 These sealing adaptors can be used for anti-biohazard.
AF-1408A
8 x 14mL
* 6
* 5
*3,4
*3
*3
*3
*5
*6
*7
Fixed Angle Rotor (for Conical Tubes)
Fixed Angle Rotor (Middle Volume)
*
Autoclavable
*
Autoclavable
*
Autoclavable
*
Autoclavable
*
Autoclavable
*
Autoclavable
AF-5008C
8 x 50mL Conical tube
*
The number of times allowed for autoclaving is limited. For more details, please contact us.
Rotor has prescribed operating life cycle, after that the rotor must be replaced.
For more details, please contact us.